Humboldt Deluxe Series Скачать руководство пользователя страница 3

Product Description

This product is intended for use only in accordance with the directions and 

specifications contained in this User Guide. While Humboldt Deluxe Water 

Baths can be used for many uses within labs, they were designed specifically 

for use in providing a consistent temperature bath for storing asphalt Marshall 

specimens in accordance with ASTM D6926, AASHTO T245 and British 

Standard BS 598: Part 107. The control processor in all our water baths provides 

a temperature range from ambient to 180°F (82°C) accurate to within 0.1% of 

input span ±1°C.

Humboldt Water Baths are fully insulated to help maintain constant 

temperatures easily. Models H-1390 and H-1392 can accommodate (12) 4" 

diameter or (3) 6" diameter Marshall specimens at a time. And the model 

H-1394 can accommodate (16) 4" and (9) 6" diameter Marshall specimens at 

a time. All models include a stainless steel shelf, which supports specimens 

while allowing 2" of free circulating water above and below specimens. Models 

H-1390 and H-1394 also utilize a magnetic stirring bar to induce water flow 

within the bath and ensure a uniform temperature is maintained. Model H-1392 

does not have a magnetic stirring bar.

Initial Installation

Your Humboldt water bath was thoroughly inspected and tested prior to being 

shipped and should be ready to operate once initial installation procedures 

are completed. To begin, please remove all packaging. Water baths are 

packed, one to a box, with the bath, shelf, stirring bar and glass stirrer plate all 

individually wrapped. Please make sure you remove the tape from the stirring 

bar and glass stirrer plate. Inspect all components and make sure they are free 

of any damage, which could have occurred during shipping. If shipping damage 

is observed, please contact Humboldt or your local agent and file a claim with 

any carriers involved.

You Humboldt water bath should be placed on a table or bench that is level. 

Check unit for power requirements and connect the line cord plug into a 

suitable electrical outlet.

Using the white stirring bar, locate the magnetic field, generated on the bottom 

of the water bath tank. Once the field in located, place the round glass stirrer 

plate over the field and place the stirring bar on top of the glass stirrer plate.

The stirring bar should spin in the middle of the glass, if not, shift the stirrer 

plate position until the stirring bar is located in the center. ( See figure #1)

Содержание Deluxe Series

Страница 1: ...product manual 05 08 H 1390 H 1392 H 1394 Humboldt Deluxe Water Baths ...

Страница 2: ...nd all exposed parts are stainless steel Front panel is water and corrosion resistant Models covered in manual include Deluxe Water Bath 110V 60Hz H 1390 Deluxe Water Bath 220V 50 60Hz H 1390 4F Microprocessor based temperature control for precise temperature control throughout the range Includes magnetic stirrer ensuring constant water tem perature and a stainless steel shelf which stands 2 51mm ...

Страница 3: ...rature is maintained Model H 1392 does not have a magnetic stirring bar Initial Installation Your Humboldt water bath was thoroughly inspected and tested prior to being shipped and should be ready to operate once initial installation procedures are completed To begin please remove all packaging Water baths are packed one to a box with the bath shelf stirring bar and glass stirrer plate all individ...

Страница 4: ...ater Bath is calibrated to a certified thermometer to ensure accurate temperatures during operation In most cases you should not need to recalibrate you water bath However if you need to access the controller for calibration purposes please contact Humboldt for step by step directions WARNING DO NOT RUN UNIT DRY UNIT SHOULD NEVER BE USED WITH LESS THAN 2 INCHES OF WATER IF UNIT IS EMPTIED WHEN HOT...

Страница 5: ...ler to set the desired temperature you would like to operate the bath at i e the setpoint The setpoint is the lower number which is in green Humboldt Deluxe Water Baths are only capable of heating the bath there is no cooling function Once you have set the setpoint for the bath the heating element will begin to heat the bath This will be indicated by a flashing light to the right of the actual tem...

Страница 6: ...ure 3 Push the blue scroll button on the far left of the controller twice which should change the controller view to this Figure 4 Now you can use the arrows to toggle between F and C Set the controller to the desired choice and push the blue scroll button repeatedly until the unit returns to the setpoint and current temperature view Once there the blue button will no longer change the display Thi...

Страница 7: ...hine Water baths for 120V service use a 15 amp Optifuse FCA 15A fuse The Humboldt replacement part for this fuse is H 2986 6 1 Water baths for 220V service use a 7 5 amp Optifuse FCA 7 5A fuse The Humboldt replacement part for this fuse is H 2986 6 2 Maintenance Water baths can generally be maintained by keeping them clean and properly filled WARNING DO NOT RUN UNIT DRY UNIT SHOULD NEVER BE USED W...

Страница 8: ...r parts of such product for the warranty of this product please refer to Humboldt Mfg Co catalog on Terms and Conditions of Sale The purchaser is responsible for the transportation charges Humboldt Mfg Co shall not be responsible under this warranty if the goods have been improperly maintained installed operated or the goods have been altered or modified so as to adversely affect the operation use...

Отзывы: