③ USE
button to edit the information of device.
④ Press
button to save the information after editing completely.
<Device Setting detail example>
-
Select Marker
Use
to select the marker between enabled and disable marker and press
button to check /uncheck.
-
Quantitative or Qualitative
Use
to select either quantitative or qualitative and press
button to check /uncheck.
(In case of selecting ‘Qualitative’, the result is showed an positive or negative by ref. range.)
-
Ref. Range
Set Reference Range for each marker. The device is set up with the default reference range.
Use the external keyboard to change from default.
(Only change it when it is necessary within the certified range.)
-
Set Performance Value
Set Correlation Function between devices to revision analysis.
Use along “Set” to select the corresponding marker and press to check/uncheck.
(Use it only when it is necessary with definite standard correlation with two markers.)
"Don’t use “DEVICE SETTINGS” menu except a set point extracted from certified standard and method.”
1. Select Marker
3. Ref. Range
2. Quantitative or Qualitative
4. Set Performance Value
Содержание HUBI-QUAN PRO
Страница 1: ...Rev 03 2016 07 28 Code No KM I 2153 ...
Страница 10: ...Close HUBI QUAN Pro System Press the Power Button Select Shut down by and press to close the system ...
Страница 26: ...9 Press to print out the test result 10 Move back to the Pat ID Screen automatically and wait for next test ...
Страница 29: ...8 Press to print out the test result 9 Move back to Pat ID Screen automatically and wait for next test ...