background image

4

number

Description

number

Description

1

AC plug

2

AC ON

3

Test Port

E

G

L

4

LCD display

5

Key

Key and function

Key numbered from left to right, the first for a total of five off down the power Key.

key

number

Description

1

Press the switch on button startup, shutdown condition

2

3

4

functional key

5

The TEST key, press and hold 1 s TEST, then stop (enter the
Settings menu is invalid)

Test setup interface

Set test voltage

SetThe test time

Battery Indicator

Set Short Current

Set the filter

Button function tip

Soft key

Description

OFF

Power Off (No operation, 3 minutes automatic shutdown)

H.V

Set test voltage: 500V

1000V

2500V

5000V

10000V

TIME

SetThe test time: 00:30 01:00

10:00

30:00

Switching software keys to a new function, or from other
function to return

I

Set Short Current

1 mA step selection Usually according to

the maximum, can also according to the regulations stipulated
by the short circuit current value choice.

Filter

Set the filter:
0

without filtering; (no interference case recommended)

1

hardware low-pass filter; (applicable to most situations)

2

software Low pass filter;High resistance (greater than 100G

recommended choose)
3

ha software low-pass filter;

(high resistance is greater than 1T recommend choose)

HELP

About Tester message

Battery Indicator

、 、 、 、

Содержание HZJY-10K-I

Страница 1: ...HZJY 10K I Insulation Resistance Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Страница 2: ...Contents I Safe Use Items 1 II Feature 2 III Specification Parameter 2 IV Introduction Of Instrument 3 V Instrument Using 6 VI Packing List 7...

Страница 3: ...operators in instrument surface operation Please don t charged the connection test line low Low in the measurement or do not touch the circuit being measured immediately after the test may result in...

Страница 4: ...al discharge when testing cables automatic rapid discharge of instruments 2 power supply modes the battery is powered by lithium ion battery and the battery life can reach 4 hours 10000V 100m test res...

Страница 5: ...range 20uF Capacitance testing accuracy 15 0 03uF Capacitor discharge rate from 10000V to 10V 0 5S F Insulation resistance testing range and accuracy T 23 5 C RH 45 75 range accuracy 500V 1000V 2500V...

Страница 6: ...Power Off No operation 3 minutes automatic shutdown H V Set test voltage 500V 1000V 2500V 5000V 10000V TIME SetThe test time 00 30 01 00 10 00 30 00 Switching software keys to a new function or from...

Страница 7: ...1 key the power button until into the contrast adjustment menu 3 Press 1 or 2 keys to adjust contrast until the right 4 Press the 4 key save and exit if you don t need to adjust the press 3 key to ex...

Страница 8: ...oltage and start 3 The instrument automatically record 15S 60S and 600S resistance and calculate the polarization index after 600 seconds Polarization index calculation method PI R600 R60 test connect...

Страница 9: ...7 VI Packing List No Item Qty 1 Main engine 1 2 Power line 1 3 Black test line 1 4 Red test line 1 5 Green test line 1...

Отзывы: