HZJY-1016 Insulation Tester
- 5 -
●
Test setup interface
SetThe test time
Battery Indicator
Set Short Current
Set the filter
Button function tip
Soft key
Description
OFF
Power Off (No operation, 3 minutes automatic shutdown)
H.V
、
1000V
、
2500V
、
5000V
、
10000V
TIME
SetThe test time: 00:30 01:00
10:00
30:00
Switching software keys to a new function, or from other function
to return
I
Set Short Current
,
1 mA step selection Usually according to the
maximum, can also according to the regulations stipulated by the
short circuit current value choice.
Filter
Set the filter:
0
:
without filtering; (no interference case recommended)
1
:
hardware low-pass filter; (applicable to most situations)
2
:
software Low pass filter;High resistance (greater than 100G
recommended choose)
3
:
ha software low-pass filter;
(high resistance is greater than 1T recommend choose)
HELP
About Tester message
Battery Indicator
:
、 、 、 、
When the battery voltage shows just one more case should be timely charge, when the
battery displaying flashing is automatically cut off the power at any time, should be
recharged.
charge
:
AC socket provided meet the requirements of the AC power supply for panel, you
can recharge equipment charging, the power indicator, display the battery label, in turn,
from
(0) to (full) showed that when filled, displays (full)
Contrast Settings interface: LCD display gray level change due to temperature changes,
may affect the electronic contrast adjustment using the instrument Settings Follow these
steps to set up:
1.
In shutdown state, hold key 2 (the 2 key).
2.
Press the 1 key (the power button), until into the contrast adjustment menu.
3.
Press 1 or 2 keys to adjust contrast, until the right .
4.
Press the 4 key save and exit, if you don't need to adjust the press 3 key to exit.