3
III. MAJOR TECHNICAL INDICATORS
Voltage supply:
AC 220V±10%
Power frequency
:
50Hz/60Hz ±1%
Measuring range
:
Electric capacity
5pF
~
200pF
Relative Permittivity
1.000
~
30.000
Dielectric loss factor
0.00001
~
100
DC Resistivity
2.5MΩm
~
20TΩm
Measuring accuracy
:
Electric capacity
± (1% of measured value+0.5pF)
Relative Permittivity
±1% of measured value
Dielectric dissipation factor
± (1% of measured value+0.0001)
DC Resistivity
±10% of measured value
Resolution
:
Electric capacity
0.01pF
Relative capacitivity
0.001
Dielectric dissipation factor
0.00001
Measured temperature range
:
0
~
125
℃
Temperature measurement error
:
±0.5
℃
Test Voltage AC(RMS)
:
500
~
2200V Continuously adjustable, frequency 50Hz
Test Voltage DC
:
0
~
500V Continuously adjustable
Power consumption
:
100W
External dimensions
:
500×360×420
Total weight
:
22Kg
Ⅳ
.USE CONDITIONS
Ambient temperature
:
0
℃~
40
℃
Relative humidity(RH)
:
<75%
Ⅴ
.PANEL DESCRIPTION AND OPERATION MATTERS
NEEDING ATTENTION
1. Panel description