If there is no sign of damage, unpack the packing case, take out the packing list, and

check whether the attached accessories are consistent with the packing list. If there is any

difference, please contact our sales department. After checking the attached accessories,

remove the main engine from the bottom plate of the box and place it on a flat ground. The

instrument installation shall meet the following requirements:

1. Environmental requirements

a. The ground shall be solid and flat concrete floor, without vibration source.

b. The highest altitude is 2000 meters.

c. This equipment is limited to indoor work. The centrifuge is suitable for constant

temperature (about 20

), the maximum relative humidity is 80% at 31

and 50% at

40

. Therefore, the centrifuge should not be placed near the heat source (such as direct

sunlight, heating pipe and radiator, etc.).

2. Installation space requirements:

The centrifuge should have enough space to ensure the air flow around the centrifuge,

otherwise it may lead to overheating and poor effect.

3. Power requirements

The power supply shall be 220V 15A single-phase power supply with independent

ground wire. It is not allowed to replace the ground wire with zero wire. The centrifuge

must be properly grounded.

4. Installation requirements of the machine: ensure safety and centrifugal effect after

being placed in place. The instrument must be placed on a solid, shockproof and

horizontal platform, and ensure that four machine feet are evenly stressed.

5. Plug in the power cord (external power separator)

VII

Operation method

1. Power on the instrument

The upper right corner of the instrument is equipped with a power switch. Move the

switch upward to the "on" end ("off" end is off), and the centrifuge is powered on.

2. Open the door cover

A "manual door opening" device is arranged at the groove on the right side of the

Содержание HZDD5Y

Страница 1: ...HZDD5Y Centrifuge Huazheng Electric Manufacturing Baoding Co Ltd...

Страница 2: ...n technical parameters 2 IV Technical parameters of rotor 3 V Shape and mechanism 3 VI Installation 3 VII Operation method 4 VIII Precautions 7 IX Maintenance 8 X Common faults and troubleshooting 8 X...

Страница 3: ...p automatically and give an alarm to ensure the safety of the instrument in use The instrument adopts the optimized design scheme for the shock absorption system The instrument operates stably with sm...

Страница 4: ...n farthest from the axis Rmin radius of rotation of the test solution nearest to the axis density of mixed liquid g cm3 viscosity of mixture poise N speed RPM R particle radius CM particle density g c...

Страница 5: ...hape and mechanism As shown in the figure VI Installation After the instrument arrives at your company check whether the appearance of the packing box is severely damaged in the process of transportat...

Страница 6: ...heating pipe and radiator etc 2 Installation space requirements The centrifuge should have enough space to ensure the air flow around the centrifuge otherwise it may lead to overheating and poor effe...

Страница 7: ...inspection if conditions permit balance weight is recommended 2 If the reagent is not enough empty test tube shall be used for water injection instead of unbalanced operation 3 Finish the work above...

Страница 8: ...e window flashes the time setting will be entered and then press the plus or minus key to determine the running time 4 When the above steps are completed press the approval key in time to confirm othe...

Страница 9: ...l sound before leaving 5 During the maintenance of the instrument the power plug shall be removed otherwise electric shock may occur easily Tips 1 The hanging cup shall be placed on the rotor hanging...

Страница 10: ...om the centrifugal chamber apply a thin layer of grease to the center hole of the rotor and the main shaft of the centrifuge and keep the rotor in a dry place X Common faults and troubleshooting Some...

Страница 11: ...t The key is damaged Replace panel Motor damage or leakage Replace It can run but can t go up in speed The machine has strange noise or peculiar smell Control system or motor failure Send to the manuf...

Страница 12: ...is mainly used to protect the semiconductor elements of the instrument control system The fuse is a closed glass tube structure and is connected by plug in type It has compact structure safe use and...

Страница 13: ...llowing situations of supporting rotors Use when the maximum speed of rotor is exceeded The rotor is scratched or bruised The rotor is seriously corroded by acid and alkali Repair by unauthorized prof...