6
3 Click the "History" button to enter the historical data query interface, which is used to query the
sample data made before.
4 Click the "Back" button to exit the interface automatically and return to step 2.
5 The "Save" button is gray, which means that pressing this button is invalid in the current state.
6 The lower left area of the interface shows the temperature of each important volume point during
the distillation.
7 The top shows the temperature-volume curve during heating, where the x-axis is the volume and
the y-axis is the temperature
8 The middle area under the interface is the real-time values of various parameters during the
system's operation, including atmospheric pressure, heating furnace temperature, steam
temperature in the distillation flask, refrigeration tube temperature, recovery chamber
temperature and recovery volume.
Note:
Before clicking "Start", you must ensure that all test parameters are set correctly. Otherwise,
once the test is started, you cannot modify the parameters again. If you click the "Force Stop"
button during the test, the test will end in advance. It means unsuccessful and all data of the test is
invalid. Click the "Back" button to restart the test.
===============================================================
5
〈
Parameter setting interface
〉
1. Click the "Hand" area to set the parameters. Enter a maximum of one decimal place. After
selecting the parameters, click the "Set" button to determine the parameters. The parameters will
be automatically saved after each modification, and CE is Grid key.