Model
Cat. no.
Volume
[μl]
Accuracy
[%]
Precision
[%]
Fit to tips
μl
Shafts color
for D
DV2
D2
4041
4051
0.2
1.0
Max
2.0
± 12.0
± 2.7
± 1.5
± 6.0
± 1.3
± 0.7
10 μl
red
DV10
D10
4042
4052
Min
0.5
5.0
Max 10.0
± 4.0
± 1.0
± 0.5
± 2.8
± 0.6
± 0.4
DV20
D20
4043
4053
Min
2
10
Max
20
± 3.0
± 1.0
± 0.8
± 1.5
± 0.5
± 0.3
200 μl
yellow
DV50
D50
4047
4057
Min
5
25
Max
50
± 2.5
± 1.0
± 0.8
± 2.0
± 0.6
± 0.4
DV100
D100
4044
4054
Min
10
50
Max
100
± 1.6
± 0.8
± 0.8
± 0.80
± 0.24
± 0.20
DV200
D200
4045
4055
Min
20
100
Max
200
± 1.2
± 0.8
± 0.6
± 0.60
± 0.25
± 0.20
DV250
D250
4040
4050
Min
50
125
Max
250
± 1.0
± 0.8
± 0.6
± 0.4
± 0.3
± 0.3
300 μl
green
DV1000
D1000
4046
4056
Min
100
500
Max 1000
± 1.6
± 0.7
± 0.6
± 0.40
± 0.20
± 0.15
1000 μl
blue
DV5000
D5000
4048
4058
Min
500
2500
Max 5000
± 1.2
± 0.6
± 0.5
± 0.50
± 0.20
± 0.15
5000 μl
white
DV10000
D10000
4049
4059
Min 1000
5000
Max10000
± 2.5
± 0.8
± 0.5
± 0.6
± 0.3
± 0.2
10000 μl
Tip ejector adjustment (Fig. 6C)
1. Remove the ejector pushbutton.
2. To increase the ejector length turn the plastic bush to
the left.
3. To decrease the ejector length turn the plastic bush to
the right.
- in 5000 & 10000 μl pipettes
Tip ejector disassembly (Fig. 7A)
1. Remove the tip ejector button
2. Use the screwdriver to turn the metal arbor left to
unscrew the ejector from the arbor.
3. Release the ejector.
Tip ejector assembly (Fig. 7A)
1. Remove the tip ejector button.
2. Use the screwdriver to turn the metal arbor right to
screw the ejector in and adjust its length.
Tip ejector adjustment (Fig. 7A)
1. Remove the ejector pushbutton.
2. Use the screwdriver to increase or decrease the ejec-
tor length.
After ejector assembly or adjustment, put the ejector but-
ton back on its place.
If above described method of ejector adjustment is not
sufficient or the diameter of the ejector opening is too
large to eject the tip it is necessary to put the ejector cap
“M” onto the ejector, (Fig. 7B).
- in 2 and 10 μl pipettes (Fig. 7C)
The 2 μl - 10 μl pipettes may require the use of an ejector
cap "M" to efficiently eject certain brands of pipette tips.
Simply place the cap, supplied with the pipette, on the
bottom of the pipette shaft and slide the cap upwards
until it surrounds the bottom of the tip ejector.
The DISCOVERY Comfort is a high quality instrument
which offers excellent accuracy and precision.
The accuracy and precision of liquid sampling depend on
the quality of tips used. The values for accuracy and preci-
sion given in the table below were obtained using HTL tips.
Only those tips guarantee correct operation of the pipettes
and ensure accuracy and precision of liquid sampling.
ENGLISH
The accuracy and precision are obtained with HTL tips,
using a gravimetric method, performing at least 10 mea-
surements of distilled water at a temperature of 20±1°C,
according to EN ISO 8655 standard.
The pipette design enables the user to perform the
recali bration process according to the rules presented in
section 8.
2 - SETTING THE VOLUME
The volume display shown by the counter has three dig-
its, which should be read from top to bottom. In addition
the lowest counter drum is printed in a scale which
enables the setting of the volume in the minimum gradua-
tion range. Examples of the meanings of the black and
red digits:
4
3
Содержание DISCOVERY Comfort D10
Страница 1: ...ENGLISH 1 16 DEUTSCH 17 34 FRANÇAIS 35 52 ESPAπOL 53 68 PORTUGU S 69 84 POLSKI 85 102 êìëëäàâ 103 120 ...
Страница 2: ...A N F G J I C A1 A2 B S C P D E D L A B C D E F ...
Страница 63: ...5 4 8 7 6 7A 5000 µl 10000 µl 7B 5000 µl 10000 µl 7C 2 µl 10 µl 6A 6B 6C S S 5A 5B 4D 4A 4B 4C ...
Страница 64: ...Daniszewska 4 03 230 Warsaw Poland Tel 48 22 492 19 00 Fax 48 22 492 19 93 Producent ...