The MAX/MIN button is used to measure the maximum and 

minimum level of sounds.Press the MAX/MIN button two 

seconds,the meter will be go on measuring.

5)On/Off button

6)Tripod hole

fixed the meter on the tripod

7)Battery cover

V.Measurement Preparaion

(1)  Battery loading

 

Remove the battery cover on the back and put in one  

 

9V Battery.

(2)  Battery Replacement

 

When the battery voltage drops below the operating  

 

voltage, mark"BAT"appears. If it appears, battery   

 

should be replaced with new one.

VI. Operating Precautions

(1)  Wind blowing across the microphone would bring  

 

additional extraneous noise.

 

Once using the instrument in the presence of wind, it is  

 

a must to mount the windscreen to not pick up    

 

undesirable signals.

(2)  To achieve more accurate measurement, use an   

 

extension cable to separate the Microphone from the  

 

main body so that the effect of unexpected sound  

 

reflection can be eliminated.

(3)  Calibrate the instrument before operation if the    

 

instrument was not in use for a long time or operated at  

 bad environment.

1) Windscreen

If you operate at wind speed over 10m/sec, please put 

protective accessories in front of the microphone.

2)Display

SYMBOL FUNCTION

LCD 

 3digits

MAX/MIN 

Maximum and Minimun Value

OVER 

Over range

dBA unit

HOLD 

Keep the reading

BAT 

 Low-Battery

3)"HOLD" button

Data Hold button: Press and hold the Button to trun on or off 

data hold function

The hold function freezes the reading in the display.

4) "MAX/MIN"button

4

5

Содержание HT-80A

Страница 1: ...by qualified personnel 2 Periodically wipe the case with a dry cloth Do not use abrasives or solvents on this instruments II General Description Thank you for using our Sound Level Meter To ensure that you can get the most from it we recommend that you read and follow the manual carefully before use This Sound Level Meter has been designed to meet the measurement requirements of safety Engineers H...

Страница 2: ... Alarm function OL is show when input is out of range Auto power off Meter automatically shuts down after approx 15 minutes of inactivity Power supply One 9V battery 006P or IEC 6F22 or NEDA 1604 Power life About 50hrs alkaline Battery Operation temperature 0 to 40 C 32 to 104 F Operation humidity 10to 90 RH Storage temperature 10 to 60 C 14 to 140 F Storage humidity 10 to 75 RH Dimensions 170 58 ...

Страница 3: ...n the presence of wind it is a must to mount the windscreen to not pick up undesirable signals 2 To achieve more accurate measurement use an extension cable to separate the Microphone from the main body so that the effect of unexpected sound reflection can be eliminated 3 Calibrate the instrument before operation if the instrument was not in use for a long time or operated at bad environment 1 Win...

Страница 4: ...d pressure level will be displayed 3 When MAX maximum mode is chosen The instrument wil show the maximum and minimum reading on the display 4 When HOLD data hold mode is chsen The hold function freezes the reading in the display Press the HOLD button momentarily to activate or to exit the HOLD function 5 Turn OFF the instrument and remove and remove battery when not in use 6 Dongguan Xintai Instru...

Отзывы: