LCD Display 

 

Power Switch 

 

Backlight set switch 

 

Data-hold Switch (H) 

 

Rotary Switch 

 

Input Jack 

 

 

Case 

 

Battery Door 

 

Brace 

 

Screw 

2.2   

FUNCTION AND RANGE SELECTOR

 

 

This meter is a portable professional measuring instrument with large LCD display and back light for easy reading. 

 

This  meter  has  the  function  of  preventing  the  test  leads  from  wrong  connecting.  The  input  socket  for  red  test  lead  is  arranged  with 

proper functions and ranges, when the transform switch can’t be rotated, stop rotating. It means the selected range isn’t suitable with 

position  of  the  red  lead  socket.  Pull  out  the  read  lead  and  then  select  the  range  required, this  provides  protection  for  meter  to  avoid 

damage by operating improperly. 

 

This meter has function of Data-hold. 

 

A rotary switch is used to select functions as well as ranges. 

3. SPECIFICATIONS 

 

Accuracy is specified for a period of year after calibration and at 18

  to 28

  (64

  to 82

) with relative humidity to 75%. 

3.1 

GENERAL SPECIFICATIONS

 

3.1.1 It includes 30 ranges of function. 

3.1.2 Over-range protection for all ranges. 

3.1.3 Operating altitude: 2000 meters (7000ft.) maximum. 

3.1.4 Max Voltage between Terminals and Earth Ground: 1000V DC or AC (peak value). 

3.1.5 Fuse protection: F1-200mMA/250V    F2-10A/250V 

3.1.6 Display: 28mm LCD, word is 28mm high 

3.1.7 Max Show Value: 1999 (31/2) 

3.1.8 Polarity Indication: “-” indicates negative polarity. 

3.1.9 Over-range Indication: display “1” or “-1”. 

3.1.10 Sampling Time: approx. 0.4 second. 

3.1.11 Power Supply: 9V battery, 6F22 or NEDA 1604. 
3.1.12 Low Battery Indication: ‘

’ displayed 

3.1.13 Temperature Factor: <0.1x Accuracy/

 

3.1.14 Operating Temperature: 0

  to 40

(32

  to 104

3.1.15 Storage Temperature:-10

  to 50

(10

  to 122

3.1.16 Dimension: 191x82x36MM (without holster) 

3.1.17 Weight: approx. 280g (including battery) 

 

3.2 

ELECTRICAL SPECIFICATIONS

 

Circumstance Temp.: 23

±

5

        Relative Humidity: <70% 

3.2.1 

DC Voltage

 

Range 

Resolution 

Accuracy 

200mV 

0.1mV 

±

(0.5% of rdg + 1 digit ) 

2V 

1mV 

±

(0.5% of rdg + 3 digits) 

20V 

10mV 

200V 

100mV 

1000V 

1V 

±

(0.8% of rdg + 3 digits) 

-- Input Impedance: 10M

Ω

 

-- Overload protection:    200mV range: 250V DC or rms AC 

                      2V-1000V ranges: 1000V DC 0r 700V rms AC 

NOTES:   

At the little voltage range, the meter will show unsteady reading when the test leads haven’t reach the circuit, it’s normal  because the meter is 

very sensitivity. When the test leads touch the circuit, you can get the true reading. 

 

3.2.2 

AC Voltage

 

Range 

Resolution 

Accuracy 

2V 

1mV 

±

(1.0% of rdg + 5 digits) 

20V 

10mV 

200V 

100mV 

700V 

1V 

±

(1.2% of rdg + 5 digits) 

-- Input Impedance: 10M

Ω

 

-- Overload protection:    200mV range: 250V DC or rms AC 

                      2V-700V ranges: 1000V DC 0r 700V rms AC 

-- Frequency Range: 40 to 400Hz 

-- Test on 60Hz/50Hz 