MULTITEST M72 - M73 - M74 - M75
EN - 27
4.8.2. Anomalous
cases
which may occur during
Ω
0.2A tests
1.
If the following condition occur:
R
MEASURED
- R
CALIBRATION
< -0.02
Ω
the instrument displays the screen beside
and emits a prolonged sound to signal
the anomalous situation
2.
If the voltage present at the terminals is
higher than 10V the instrument does not
perform the test and emits a prolonged
sound to signal the anomalous situation.
The screen beside is displayed for 5
seconds, after which the instrument
displays the default value related to the
Ω
0.2A test
Input voltage
3.
If the resistance value is higher than the
full scale the instrument emits a
prolonged sound to signal the anomalous
situation. A screen similar to this is
displayed. The same message may also
mean that mesuring cables are
disconnected or open