HT9014 - HT9015
EN - 12
4.3.5. Capacitance measurement
CAUTION
When testing in-circuit capacitors, turn off the power of the circuit to be tested
and discharge all the capacitors
Fig. 7: Taking Capacitance measurement
1. Rotate the switch on
Ω
CAP
position
2. Pushing
MODE
key and select capacitance test. The “
nF
”
symbol is shown at display
3. Insert the red test lead plug into
V
Ω
CAPHz%Temp
jack and the black test lead
plug into
COM
jack (Fig. 7)
4. Connect the two long ends of test leads to the desired circuit, then reading will be
displayed. Bargraph is disabled in capacitance measurement
5. When “
O.L.
” symbol is displayed, the result under test exceeds the maximum value that
the instrument is capable of measuring
6. For HOLD and REL features please refer to § 4.2
CAUTION
For readings of capacitance <40nF push
PEAK/REL
key before making test in
order to improve the measured result