HT7051 - HT7051N
EN - 9
5.2 INSULATION
MEASUREMENT IN FIX MODE
This measurement mode, performed in compliance with IEC/ EN61557-2 guideline, allows
a quick selection of the test voltage among the values: 250, 500, 1000, 2500, and
5000VDC.
CAUTION
The instrument may be used on installations of overvoltage category CAT
IV 600V to earth with maximum voltage 600V between inputs. Do not
connect the instrument to installations with voltages exceeding the limit
values indicated in this manual. If these limit values are exceeded, the
operator may be exposed to electrical shocks and the instrument could get
damaged
Always connect the measuring leads to the instrument and to the alligator
clips with the accessories disconnected from the system
It is recommended to hold the alligator clip respecting the safety area
marked by the hand protection (see § 4.4)
Fig. 4: Instrument connection
1. Turn on the instrument by press
ON/OFF/
key
2. Press the
FUNC/ERASE MEM
key to select the “FIX” mode
3. The display shows a screen similar to
that reported here to the side, which
contains:
The FIX selected mode
The bargraph set to zero
The "--- T
Ω
" indication
The selected test voltage
The test time value
The plug symbol (if the instrument
is supplied from the mains)
The battery symbol steadily
illuminated or flashing (if batteries
are flat or are being recharged)
The TIMER symbol if the duration
of the test has been set