Temperature with K probe
Range Resolution Accuracy
Overload
protection
-200.0°C ÷ 0.0°C
0.1°C
(1.0%rdg+2°C)
1000VDC/ACrms
0.0°C ÷ 1200.0°C
(1.0%rdg+1°C)
-328.0°F ÷ 32.0°F
0.1°F
(1.0%rdg+36°F)
32.0°F ÷ 2192.0°F
(1.0%rdg+18°F)
Insulation resistance
Test voltage
Measuring range
Accuracy Overload
protection
50V DC
2.000M
(1.5%rdg+5dgt)
600VDC/ACrms
20.00M
55.0M
100V DC
2.000M
20.00M
110.0M
250V DC
2.000M
20.00M
200.0M
275M
500V DC
2.000M
20.00M
200.0M
550M
1000V DC
2.000M
20.00M
200.0M
2000M
22.0G
(10%rdg+3dgt)
Test voltage accuracy: +20%rdg, -0%rdg
Short-circuit current: 1mA
Minimum resistance (@ nominal current 1mA): 50k
(50V), 100k
(100V), 250k
(250V), 500k
(500V), 1M
(1000V)
Discharge time of measured object: <1s (C
1µF)
Maximum capacitive load: 1µF
Voltage detection on circuit: test inhibited for voltages
30V AC/DC on inputs