5.3. FUSE REPLACEMENT
CAUTION
Before replacing fuses, disconnect test leads from any energized circuit to
avoid electrical shock.
LEGEND:
1. Turn
OFF
the
meter
and
disconnect the test leads from the
input terminals.
2. Remove the protective holster from
the meter.
3. Unscrew the battery cover and
remove the battery.
4.
Remove the defective fuse and
install a new fuse of the same size
and rating (fast type 10A/1000V).
Make sure the new fuse is
centered in the fuse holder. Re-
screw the battery cover and
reposition the protective holster.
Fig. 11: Fuse replacement
5.4. CLEANING
To clean the instrument use a soft dry cloth. Never use a wet cloth, solvents or water.
5.5. END OF LIFE
CAUTION
: this symbol indicates that equipment and its accessories shall be
subject to a separate collection and correct disposal.