9. TECHNICAL SPECIFICATIONS
9.1. TECHNICAL FEATURES
Accuracy indicated as [%r (number dgts * resolution)] at 23 ± 5°C, <80%RH
3- and 2-wire earth resistance measurement - EARTH 3W and EARTH 2W
Measurement range (**)
Resolution [
]
Accuracy (*)
Reading [
] Measure
[
]
0.01
19.99
0.08
19.99
0.01
(2.5%rdg + 2dgt)
20.0
199.9
20.0
199.9
0.1
200
1999
200
1999
1
2.00
19.99k
2.00
19.99k
0.01k
20.0
49.9k
20.0
49.9k
0.1k
Ground resistivity measurement -
(distance among the rods=1m)
Range (**)
Resolution [
m]
Accuracy (*)
Reading [
m] Measure
m]
0.06
19.99
0.50
19.99
0.01
(2.5%rdg+2dgt)
20.0
199.9
20.0
199.9
0.1
200
1999
200
1999
1
2.00
19.99k
2.00
19.99k
0.01k
20.0
199.9k
20.0
199.9k
0.1k
200
314k
200
314k
1k
Measuring frequency: 77.5Hz; Test current:
12mA; Open loop voltage: < 25Vrms
Disturbance voltage on ampere and volt circuits: the measurement is taken with the stated accuracy if the interfering
voltage is
3V, while for interfering voltages ranging from > 3V and
9V, the accuracy decreases progressively;
with an interfering voltage equal to 9V the instrument does not perform the test
(*) If
R
P
> 1200 + 100 R
X
and/or R
C
> 1200 + 100 R
X
, R
P
> 50k
and/or R
C
> 50k
and the instrument carries out the
test, the accuracy of the instrument is
(10% reading) where:
R
P
= resistance of the voltage circuit; R
C
= resistance of the current circuit; R
E
= measured earth resistance
(**) Automatic selection of the range
Interfering voltage measurement
Range (**)
Resolution [V]
Uncertainty
Reading [V]
Measure [V]
0
460
7
460
1
(2.0%rdg+2dgt)
(**) Automatic selection of the range