7.1.2. AIR, RH, TMP °F, TMP °C, Lux mode
3. Insert the connector of the probe being used into the instrument input In1 making sure
that the probe connected and the set measurement unit correspond
4.
Press the
GO/STOP
key. The instrument stops
updating the measured value and
STOP
appears on the right bottom part of the display.
Press the same key again to restart the
measurement and the real-time display of the
instant value of the input parameter. In this
case,
RUN
appears on the right bottom part of
the display.
A U X
I n 1 =
7 . 0 8 L u x
Lux LX
20
STOP
Func
FS
5.
Measures can be saved, both in RUN and in STOP mode, by pressing the
SAVE
key twice or the
SAVE
key and, subsequently, the
ENTER
key (§ 9.1)
7.1.3. Description of anomalous results
1. In AIR, RH, TMP °F, TMP
°C, Lux or VOLT mode, if the
instrument measures an
input value higher than the
full scale, it displays a screen
similar to the one reported
here to the side. Check that
the full scale selected on the
instrument matches the full
scale on the transducer
A U X
I n 1 > 2 0 . 0 0 k L u x
Measured value higher than the
instrument’s full scale
Lux Lx
20k
Func
FS
2.
The results displayed can be saved by pressing the
SAVE
key twice or the
SAVE
key and, subsequently, the
ENTER
key (§ 9.1)
3. “DB” sound mode only: when
starting recording without the
sound level probe being
connected to the instrument,
a screen similar to the one
on the side will be shown
A U X
S P L - - - -
d B
P e a k - - - - d B
D u r a t a 0 0 0 0 : 0 0 : 0 0
NO HT55
The instrument does not find the
sound level transducer HT55
dB
Func
4.
The previous anomalous result cannot be saved