HT Italia HT8100 Скачать руководство пользователя страница 18

   HT8100 

 

 

EN - 17

4.4.7.  DC current generation 

 

 

CAUTION 

 

 

 

The maximum output DC current generated by the instrument is 24mA with 
internal battery voltage > 4.5VDC. 

 

 

Fig. 8: Use of the instrument for DC current generation 

1. Switch on the instrument by pressing and holding the 

RANGE

 key to select the 

measuring range 

0-20mA

 or 

4-20mA

2. Select the position 

mA

 in case of programmable DC current generation or position 

mA

 for DC current generation with automatic ramp. 

3.  Use the 4-arrow selector on the front panel to select the options “%STEP”, “100%”, 

“50%”, “0%” or “END” flashing on the display and confirm with the 

ENTER

 key in case 

of selectable current generation, or press the 

MODE

 key to select the ramp type (see 

§ 4.3.7) 

4.  Insert the red cable into input lead 

A

 and the black cable into input lead 

mA

. The 

instrument automatically generates the output current considering the selected options. 
Press the 

A-HOLD

 key to suspend/restore the generation 

5.  Position the red lead and the black lead respectively in the points with positive and 

negative potential of the passive external device which must be supplied (see Fig. 8) 

6.  Turn the rotary switch to exit the function and stop generation. Remove the cable from 

lead 

A

 before turning the rotary switch. 

Содержание HT8100

Страница 1: ...Copyright HT ITALIA 2013 Release EN 1 05 08 07 2013 ENGLISH User Manual...

Страница 2: ...RECALL modes 9 4 3 6 Activation deactivation of the internal functions 9 4 3 7 Output DC current generation 10 4 3 8 Loop Power and HART 250 modes 10 4 4 Measuring operations 11 4 4 1 DC Voltage measu...

Страница 3: ...gerous voltage 30V electrical shock hazard Double insulated meter AC voltage or current DC voltage or current Connection to earth 1 1 PRELIMINARY INSTRUCTIONS This instrument has been designed for use...

Страница 4: ...alled overvoltage category is 6 7 4 Measured circuits reads OMISSIS Circuits are divided into the following measurement categories Measurement category IV is for measurements performed at the source o...

Страница 5: ...TRMS True Root Mean Square VALUE meters instruments measuring the TRMS value of the quantity being tested With a perfectly sinusoidal wave the two families of instruments provide identical results Wi...

Страница 6: ...please follow the instructions given in 7 3 2 INSTRUMENT POWER SUPPLY The instrument is supplied with four 1 5V AA IEC LR6 alkaline batteries included in the package In order to prevent battery disch...

Страница 7: ...IONS 4 1 INSTRUMENT DESCRIPTION CAPTION 1 Auto Backlight 2 LCD display 3 Function keys 4 MODE key 5 Rotary selector switch 6 Input terminal COM 7 Input terminal mA 8 Input terminal HzV 9 Input termina...

Страница 8: ...t the function 4 2 2 ENTER and CANCEL keys Pressing the ENTER key allows activating a function flashing on the display selected by means of the 4 arrow selector located on the front panel of the instr...

Страница 9: ...or current The message AUTOTEST is shown on the display This is the mode set whenever the instrument is switched on Manual Test it allows manually setting the AC or DC measurement of voltage or curre...

Страница 10: ...itch off the instrument then switch on the instrument again while pressing and holding the CANCEL key to delete the internal memory 4 3 6 Activation deactivation of the internal functions The followin...

Страница 11: ...utput current adjustment is possible with the options STEP setting of values 0 25 50 75 100 120 125 of the selected range Quick adjustment setting of values 0 50 100 of the selected range FINE setting...

Страница 12: ...se Autorange selection If voltage value is unknown select the highest range 4 Insert the red cable into input lead HzV and the black cable into input lead COM 5 Position the red lead and the black lea...

Страница 13: ...RANGE key for manually selecting the measuring range see 4 2 4 or use Autorange selection If voltage value is unknown select the highest range 4 Insert the red cable into input lead HzV and the black...

Страница 14: ...the red cable into input lead mA or A and the black cable into input lead COM 5 Connect the red lead and the black lead in series to the circuit whose current you want to measure respecting polarity a...

Страница 15: ...easurement see 4 3 3 or press the HFR key for HFR measurement see 4 3 4 4 Insert the red cable into input lead mA or A and the black cable into input lead COM 5 Connect the red lead and the black lead...

Страница 16: ...If the value of resistance is unknown select the highest range 3 Insert the red cable into input lead HzV and the black cable into input lead COM 4 Position the test leads in the desired points of th...

Страница 17: ...the leads at the ends of the diode to be tested respecting the indicated polarity see Fig 7 The value of directly polarized threshold voltage is shown on the display For a good P N junction the instru...

Страница 18: ...e options STEP 100 50 0 or END flashing on the display and confirm with the ENTER key in case of selectable current generation or press the MODE key to select the ramp type see 4 3 7 4 Insert the red...

Страница 19: ...r on the front panel to select the options STEP 100 50 0 or END flashing on the display and confirm with the ENTER key in case of selectable current generation or press the MODE key to select the ramp...

Страница 20: ...250 HART mode see 4 3 8 The message 250 HART is shown on the display 3 Insert the red cable into input lead A and the black cable into input lead mA 4 Connect the red lead and the black lead to the e...

Страница 21: ...nt 1 Remove the test leads 2 Remove the fastening screw and remove the battery compartment cover 3 Remove the batteries and insert the same number of batteries of the same type see 6 1 3 respecting th...

Страница 22: ...this range 40Hz 1kHz For non sinusoidal voltages consider the following crest factors CF 1 4 FC 2 0 Add 1 0 reading to accuracy 2 0 FC 2 5 Add 2 5 reading to accuracy 2 5 FC 3 0 Add 4 0 reading to acc...

Страница 23: ...e 2 000V 1 0 rdg 10dgt 1mA circa 3V 1000VDC ACrms AC Voltage and AC Current Frequency Range Resolution Accuracy Protection against overcharge 500 00Hz 0 01Hz 3dgt 1000VDC ACrms max 440mA 5 0000kHz 0 0...

Страница 24: ...rature coefficient 0 1x accuracy C 18 C or 28 C NMRR Normal Mode Rejection Ratio 50dB for DC quantities and 50 60Hz CMRR Common Mode Rejection Ratio 100dB from DC and 50 60Hz DCV 60dB from DC and 50 6...

Страница 25: ...122 F 45 RH Storage temperature 20 60 C 4 140 F batteries not inserted Storage humidity 80 RH This instrument satisfies the requirements of Low Voltage Directive 2006 95 EC LVD and of EMC Directive 2...

Страница 26: ...uence of improper packaging Repairs which may become necessary as a consequence of interventions performed by unauthorized personnel Modifications to the instrument performed without the manufacturer...

Отзывы: