HT3305
EN - 8
5.2.
K-TYPE TEMPERATURE MEASUREMENT
CAUTION
Do not compare the infrared temperature measurement with the
measurements made with K-type probe, since (because of the totally
different nature of the two methods) the obtained values may differ
remarkably
Measurement with the K-type probe can be used in situations where IR
measurement cannot be used (e.g.: measurement of shiny/bright surfaces
such as glass and plexiglas)
1. Open the protection cap of K-type input terminal (see Fig. 1 – part 8) and connect the
K-type probe respecting the indicated polarities (see Fig. 5)
Fig. 5: Connection of K-type probe in the instrument
2. Switch on the instrument by pressing trigger key
T
and verify the “K-TYPE” message in
the top part of the secondary display
3. Connect the probe to the device under test
4. Press and hold Trigger key
T
to activate the test
5. Release Trigger key
T
to stop measuring and freeze the last read value on the
secondary display. The message “HOLD” is shown
6. The instrument will automatically switch off after 10 seconds of idleness