Field
Wiring
31
Chapter 2
Field Wiring
About This Chapter
This chapter shows how to plan and connect field wiring to the HP E1415’s
Terminal Module. The chapter explains proper connection of analog signals
to the HP E1415, both two-wire voltage type and four-wire resistance type
measurements. Connections for other measurement types (e.g., strain using
the Bridge Completion SCPs) refer to specific SCP manual in the "SCP
Manuals" section. Chapter contents include:
•
Planning Your Wiring Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
•
Terminal Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
•
Reference Temperature Sensing with the HP E1415 . . . . . . . . . . . 37
•
Preferred Measurement Connections . . . . . . . . . . . . . . . . . . . . . . . 39
•
Connecting the On-board Thermistor . . . . . . . . . . . . . . . . . . . . . . . 42
•
Wiring and Attaching the Terminal Module . . . . . . . . . . . . . . . . . . 43
•
Attaching/Removing the HP E1415 Terminal Module . . . . . . . . . . 45
•
Adding Components to the Terminal Module. . . . . . . . . . . . . . . . . 47
•
Terminal Module Wiring Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
•
Terminal Module Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
•
Faceplate Connector Pin-Signal Lists . . . . . . . . . . . . . . . . . . . . . . . 53
Planning Your Wiring Layout
The first point to understand is that the HP E1415 makes no assumptions
about the relationship between Signal Conditioning Plug-on (SCP) function
and the position in the HP E1415 that it can occupy. You can put any type
of SCP into any SCP position. There are, however, some factors you should
consider when planning what mix of SCPs should be installed in each of your
HP E1415s. The following discussions will help you understand these
factors.
SCP Positions and
Channel Numbers
The HP E1415 has a fixed relationship between Signal Conditioning Plug-on
positions and the channels they connect to. Each of the eight SCP positions
can connect to eight channels. Figure 2-1 shows the channel number to SCP
relationship.
Содержание VXI 75000 C Series
Страница 2: ......
Страница 16: ...16 ...
Страница 18: ......
Страница 30: ...30 Getting Started Chapter 1 Notes ...
Страница 32: ...32 Field Wiring Chapter 2 Figure 2 1 Channel Numbers at SCP Positions ...
Страница 44: ...44 Field Wiring Chapter 2 Figure 2 11 HP E1415 Terminal Module ...
Страница 54: ...54 Field Wiring Chapter 2 Notes ...
Страница 61: ...Programming the HP E1415 for PID Control 61 Chapter 3 Programming Overview Diagram ...
Страница 124: ...124 Creating and Running Custom Algorithms Chapter 4 Figure 4 2 Algorithm Operating Sequence Diagram ...
Страница 136: ...136 Creating and Running Custom Algorithms Chapter 4 Notes ...
Страница 152: ...152 Algorithm Language Reference Chapter 5 Notes ...
Страница 304: ...304 HP E1415 Command Reference Chapter 6 Command Quick Reference Notes ...
Страница 308: ...308 Specifications Appendix A Thermocouple Type E 200 800C SCPs HP E1501 02 03 ...
Страница 309: ...Specifications 309 Appendix A Thermocouple Type E 200 800C SCPs HP E1508 09 ...
Страница 310: ...310 Specifications Appendix A Thermocouple Type E 0 800C SCPs HP E1501 02 03 ...
Страница 311: ...Specifications 311 Appendix A Thermocouple Type E 0 800C SCPs HP E1509 09 ...
Страница 312: ...312 Specifications Appendix A Thermocouple Type E Extended SCPs HP E1501 02 03 ...
Страница 313: ...Specifications 313 Appendix A Thermocouple Type E Extended SCPs HP E1508 09 ...
Страница 314: ...314 Specifications Appendix A Thermocouple Type J SCPs HP E1501 02 03 ...
Страница 315: ...Specifications 315 Appendix A Thermocouple Type J SCPs HP E1508 09 ...
Страница 316: ...316 Specifications Appendix A Thermocouple Type K SCPs HP E1501 02 03 ...
Страница 317: ...Specifications 317 Appendix A Thermocouple Type R SCPs HP E1501 02 03 ...
Страница 318: ...318 Specifications Appendix A Thermocouple Type R SCPs HP E1508 09 ...
Страница 319: ...Specifications 319 Appendix A Thermocouple Type S SCPs HP E1501 02 03 ...
Страница 320: ...320 Specifications Appendix A Thermocouple Type S SCPs HP E1508 09 ...
Страница 321: ...Specifications 321 Appendix A Thermocouple Type T SCPs HP E1501 02 03 ...
Страница 322: ...322 Specifications Appendix A Thermocouple Type T SCPs HP E1508 09 ...
Страница 323: ...Specifications 323 Appendix A 5K Thermistor Reference SCPs HP E1501 02 03 ...
Страница 324: ...324 Specifications Appendix A 5K Thermistor Reference SCPs HP E1508 09 ...
Страница 325: ...Specifications 325 Appendix A RTD Reference SCPs HP E1501 02 03 ...
Страница 326: ...326 Specifications Appendix A RTD SCPs HP E1501 02 03 ...
Страница 327: ...Specifications 327 Appendix A RTD SCPs HP E1508 09 ...
Страница 328: ...328 Specifications Appendix A 2250 Thermistor SCPs HP E1501 02 03 ...
Страница 329: ...Specifications 329 Appendix A 2250 Thermistor SCPs HP E1508 09 ...
Страница 330: ...330 Specifications Appendix A 5K Thermistor SCPs HP E1501 02 03 ...
Страница 331: ...Specifications 331 Appendix A 5K Thermistor SCPs HP E1508 09 ...
Страница 332: ...332 Specifications Appendix A 10K Thermistor SCPs HP E1501 02 03 ...
Страница 333: ...Specifications 333 Appendix A 10K Thermistor SCPs HP E1508 09 ...
Страница 334: ...334 Specifications Appendix A Notes ...
Страница 346: ...346 Glossary Appendix C Notes ...
Страница 388: ...388 Generating User Defined Functions Appendix F Notes ...
Страница 438: ...438 Index Writing the algorithm 129 values to CVT elements 120 values to the FIFO 121 Z ZERO CALibration ZERO 194 ...