Advanced Topics
34
h
3.
Figure 4-12 shows the calibration screen, with its options for 1- or 2-point
calibration. Note that the meter continues to show the live streaming of
sensor values. Press
%1$
to enter the correction for the current reading. The
current reading is still
6.408
, but it should read
-6.433
. Enter the correct
value as shown in Figure 4-13 and press
$OK#
to return to the calibration
screen.
4.
Note that the calibration change is not yet effective, as you have the option
of entering a second correction. The white square in the
%$1!!qqq!!!
key in
Figure 4-14 indicates that the first point correction has been logged.
5.
Press
$2$
to enter a second point correction, or press
$OK#
to make the 1-
point calibration effective (Figure 4-15). Of course, you can press
CANCL
to
abort the calibration process.
Figure 4-10
Figure 4-11
Figure 4-12
Figure 4-13
Figure 4-14
Figure 4-15