Instrument
Specication
Specications
are
the
performance
characteristics
of
the
instrument
that
are
certied.
These
specications
,
listed
in
T
able
C-1,
are
the
performance
standards
or
limits
against
that
the
HP
81554SM
can
be
tested.
Appendix
C
also
lists
some
supplemental
characteristics
of
the
HP
81554SM
and
should
be
considered
as
additional
information.
Any
changes
in
the
specications
due
to
manufacturing
changes
,
design,
or
traceability
to
the
National
Bureau
of
Standards
will
be
covered
in
a
manual
change
supplement
or
revised
manual.
The
specications
listed
in
such
a
change
supercede
any
previously
published.
P
erformance
T
ests
The
performance
tests
given
in
this
section
are
separated
into
Output
P
ower
T
est
and
Stability
T
est.
P
erform
each
step
in
the
tests
in
the
order
thay
are
given,
using
the
corresponding
test
equipment.
Note
Make
sure
that
all
optical
connections
in
the
test
setups
given
in
the
procedure
are
dry
and
clean.
The
ambient
temperature
must
be
constant
(62
C)
during
the
test.
The
power
meter
must
be
zeroed
before
the
measurement.
Output
P
ower
and
Stability
T
ests
Specications:
Optical
output
power
of
10/125m
ber:
>-1dBm
Stability
over
15
minutes
at
constant
temperature:
60.005dB
Stability
over
6h
and
61
C
env
.
temp
.
window:
60.05dB
Note
The
full
sequence
of
tests
should
be
performed
rst
at
1310nm
and
again
at
1550nm.
P
erforming
all
the
tests
for
one
wavelength
together
provides
the
warm-up
time
for
the
stability
test.
HP
81554SM
P
erformance
T
ests
D-3