5
Attach the fully-connected probe adapter/Instrumented Mobile
Module assembly to the connectors on the mother board.
Connecting to the Pentium II Instrumented Mobile Module target system
6
Adjust the screws until they are making contact with the mother
board. Ensure that the probe adapter is firmly seated in the 190-pin
Mictor connector, and that the probe adapter is parallel within
±
4
°
with the mother board (see below).
C A U T I O N
The adjustment screws must be properly adjusted so that they are touching
the mother board, and the probe adapter is parallel within
±
4
°
with the
mother board. If the probe adapter is not supported by the adjustment
screws, or if the probe adapter is not seated so that it is parallel within
±
4
°
with the mother board, damage may occur to the Mictor connectors.
Required alignment for probe adapter and target system
To connect to the target system
7