/Clean the dispenser
HP D300 Digital Dispenser
10
3 Clean the dispenser
WARNING:
Observe the following cautions when cleaning the dispenser:
Do not spray cleaning fluid directly onto the Dispenser when a Dispensehead
Cassette is installed. Always avoid exposing unused Dispenseheads to cleaning
fluids.
Do not spray or wipe the ionizer bar that is mounted under the deck between the
plate opening and the Dispenseheads.
Do not spray or wipe the sensors under the Dispenser deck. Doing so could cause
the sensors to become misaligned.
Do not wipe the pogo pins. See
Figure 3-2
Pogo pin region of pogo block which should not be sprayed or wiped with cleaning fluid
As with any lab instrument, it is a best practice to routinely clean areas of the Dispenser that come
into contact with chemicals.
1.
Turn off the dispenser as instructed in the ―
Turn off the HP D300 Digital Dispenser
‖ section
of this document.
2.
Remove any unused Dispensehead Cassettes from the Dispenser.
CAUTION:
Exposing unused Dispensehead Cassettes to cleaning fluids may compromise their
performance.
3.
Spray a cleaning solution, such as a bleach or a 70% ethanol solution, onto a fabric Texwipe or
other cloth that does not generate particles.
4.
Wipe down all accessible Dispenser surfaces, including the following (See the
― section for the location of described areas):
○
The top and bottom surfaces of the dispenser deck, avoiding the ionizer bar in the cutout
on the right side of the bottom of the deck and the sensors at the front corners of the deck
○
The pogo mechanism housing, excluding the pogo pin area as described above
○
The destination plate holder
○
The source plate holder
○
All of the outside surfaces of the dispenser
Содержание Compaq d300 Series
Страница 1: ...HP D300 Digital Dispenser Operating Manual ...
Страница 72: ...HP D300 Digital Dispenser 72 Appendix C Quick Start Guide ...