7-11
Reference
Measurement Repeatability
Measurement Repeatability
This section compares the measurement repeatability of HP 86037B with and
without the multi-wavelength meter. The measurements were performed on
the same spool of 22-km dispersion-shifted fiber. The wavelength range was
from 1525 nm to 1575 nm in 1 qnm steps. The data were analyzed with the
five-term Sellmeier fit.
The standard deviation for
λ
0
is actually higher for the measurements taken
with the multi-wavelength meter than for those without. This is most likely
because the measurements made with the multi-wavelength meter take
longer, and therefore are more susceptible to fiber length drifts. The particular
HP tunable laser source used in the measurement was so accurate that it did
not introduce much wavelength error. These results are based on one system,
therefore we cannot guarantee them as specifications for all systems since the
absolute wavelength accuracy of HP’s tunable laser sources is specified to be
±0.1 nm. For best accuracy, a system with a multi-wavelength meter is still
recommended.
Table 7-2. Repeatability Data for 20 Runs With the Multi-Wavelength Meter
Parameter
λλλλ
0
(nm)
S
0
[ps/ (nm
2
××××
km)]
Mean 1550.8938
0.0733825
Maximum 1550.9062
0.0734366
Minimum 1550.8829
0.0733237
Max-min 0.0233 0.0001129
Standard Deviation
0.008116
0.0000318
Содержание 86037B
Страница 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Страница 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Страница 72: ...3 16 Normalization Setting Normalization Preferences ...
Страница 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Страница 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Страница 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Страница 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Страница 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Страница 120: ...4 48 Remote Control Output Parameters ...
Страница 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Страница 165: ...6 7 Installation Installation 2 3 4 5 1 8 7 Fan Pack 6 6 Power Distribution Assembly 1 7 Power Distribution Assembly 2 ...
Страница 170: ...6 12 Installation Installation ...
Страница 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Страница 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Страница 190: ...7 20 Reference System Options ...
Страница 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Страница 198: ...8 8 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 204: ...Index 6 Index ...