![HP 85325A Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/hp/85325a/85325a_operating-and-service-manual_161508040.webp)
1-24
HP 85325A Operating and Service Manual
Installation
Final Installation
Proper Torque for
Coax Connections
The coax RF connectors in the setup must be properly torqued or the system
will not provide its best performance. Proper connection torque also keeps
moisture out the connectors, and eliminates radio frequency interference
(RFI) from the connector interface. The amount of torque required depends
on the type of connector. Table 1-5 lists the proper torque values. Do not
over-tighten connectors.
Turn ON AC Power
Turn ON the AC power for all of the components in the following order:
1. RF and LO sources, RF power amplifier
2. HP 85309A LO/IF unit
3. Receiver
Table 1-5
Proper Connector Torque
Connector
Torque / cm-kg Torque / N-cm
Torque / in-lbs
Wrench Part
Number
Type-N
52
508
45
8710-1935
3.5 mm
9.2
90
8
8720-1765
SMA
5.7
56
5
8710-1582
2.4 mm
9.2
90
8
8720-1765
Содержание 85325A