![HP 85106D Скачать руководство пользователя страница 112](http://html.mh-extra.com/html/hp/85106d/85106d_manual_161501112.webp)
HP 85106D mm-Wave System
4-19
Service and Troubleshooting
Check Unratioed Power Levels
Check Unratioed Power Levels
This procedure allows you to check the output power level of each
sampler/mixer assembly, individually, and each associated IF amplifier.
The typical power level display (S11 as an example) is a ratio (in this case,
b
1
/a
1
). The network analyzer automatically applies power and phase locks to
a predefined port or ports to make the selected measurement. Ratioed
measurements provide useful data but they can mask certain malfunctions.
Assume, for example, that you need to measure the S-parameter at a specific
power level of a device. If the test set has a 20 dB power hole due to a faulty
RF input connector, that deficiency would be invisible (ratioed out) in a
ratioed measurement, however the data would be incorrect. It would not
have been taken at the specified power level. Similarly, troubleshooting
system faults using the ratioed measurement mode can be deceptive.
The solution is to separately test each channel and check the power in
unratioed mode. To do so requires specifying which port receives the driven
power, and which channel is phaselocked.
The following procedure includes steps to redefine parameters as required.
The power levels given are approximate. The assemblies that are parts of the
signal path of each channel are illustrated in Figure 4-7. Knowing which
assemblies are common to two, or all four channels, is a powerful
troubleshooting tool.
Figure 4-7
Simplified Signal Path of Unratioed Power Test
Содержание 85106D
Страница 11: ...x HP 85106D mm Wave System ...
Страница 95: ...4 2 HP 85106D mm Wave System Service and Troubleshooting Figure 4 1 HP 85106D System Block Diagram ...
Страница 96: ...HP 85106D mm Wave System 4 3 Service and Troubleshooting Figure 4 2 HP 85106D Option 001 System Block Diagram ...
Страница 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Страница 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 149: ...5 36 HP 85106D mm Wave System HP 85104A Test Set Module Assembly Procedure Figure 5 12 HP 85104A Assembly Procedure ...
Страница 153: ...5 40 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 13 HP Q85104A Replaceable Parts ...
Страница 156: ...HP 85106D mm Wave System 5 43 HP 85104A Test Set Module Replaceable Parts Figure 5 14 HP U85104A Replaceable Parts ...
Страница 159: ...5 46 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 15 HP V85104A Replaceable Parts ...
Страница 162: ...HP 85106D mm Wave System 5 49 HP 85104A Test Set Module Replaceable Parts Figure 5 16 HP W85104A Replaceable Parts ...
Страница 208: ......
Страница 212: ...6 48 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 28 HP 85105A Major Assemblies ...
Страница 214: ...6 50 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 29 HP 85105A Semi rigid RF Cables ...
Страница 218: ...6 54 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 31 Wire Harnesses ...
Страница 220: ...6 56 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 32 HP 85105A Front Panel ...
Страница 222: ...6 58 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 33 Rear Panel Standard Option 050 ...
Страница 228: ...6 64 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 37 HP 85105A Accessories Standard ...
Страница 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Страница 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Страница 244: ... vi HP 85106D mm Wave System ...