D
a.
Press
4
W
avelength
5 ,
press
4
Edit
5 ,
type
1470.000
on
the
numeric
keys
,
press
4
Enter
5 .
b.
Press
4
Output
Pow
er
5 ,
press
4
Edit
5 ,
type
100
on
the
numeric
keys
,
press
4
Enter
5 .
c.
A
ctivate
the
8168A.
5.
Set
up
the
8169A.
a.
Set
8169A
to
P
ol=0
(home
position):
press
4
Home
5 .
b.
Select
the
\Circle"
application:
press
4
Circle
5 .
6.
Optimize
transmission
through
8169A:
a.
Set
the
polarizing
lter
for
maximum
transmission
(see
\Setting
the
P
osition
of
the
P
olarizing
Filter"
in
Chapter
2).
b.
On
8169A
set
to
2"
B
and
get
maximum
displayed
power
on
8153A
by
turning
knob
.
c.
On
8169A
set
to
2
P
and
get
maximum
displayed
power
on
8153A
by
turning
knob
.
Repeat
these
two
steps
until
absolute
maximum
on
8153A
display
is
reached
Note
As
the
display
shows
negative
values
,
the
maximum
displayed
power
is
the
smallest
number
displayed
7.
Note
the
displayed
value
on
8153A
in
your
test
record
in
the
column
\P
ower
after
DUT"
for
the
associated
wavelength.
8.
Repeat
list
item
6
to
list
item
7
for
the
wavelengths
between
1480
to
1570
nm
in
steps
of
10
nm,
always
setting
the
8168A
and
the
8153A
to
the
required
wavelength
for
each
setting.
9.
Repeat
list
item
6
to
list
item
7
again
for
1470
nm
to
ensure
stability
of
measurement
setup
.
If
your
measured
value
is
more
than
.01dB
o
the
previous
value
,
you
need
to
x
your
setup
and
repeat
list
item
3
to
list
item
8 .
10.
Connect
the
8168A's
output
to
the
81524A
Optical
Head
as
shown
in
Figure
D-2
.
D-6
P
erformance
T
est