Specifications
These specifications for the HP E1509 reflect the combined performance of
the HP E1413/E1313 and the HP E1509 Signal Conditioning Plug-on.
These specifications are not to be added to those presented in the
HP E1413/E1313 User’s Manual.
General Specifications
Measurement ranges
DC Volts
±
3.9mV to
±
.25V FS
Temperature
Thermocouples - -200 to +1700
°
C
Thermistors - (Opt 15 required) -80 to +160
°
C
RTD’s - (Opt 15 required) -200 to +850
°
C
Resistance
(Opt 15 required) 128 ohms to 131 Kohms FS
Strain
25,000
µ
e or limit of linear range of strain gage
Maximum input voltage
(Normal mode plus common mode)
Operating: <
±
16 V peak Damage level: >
±
42 V peak
Maximum common mode
voltage
Operating: <
±
16 V peak Damage level: >
±
42 V peak
Normal mode rejection
@ 7Hz -3dB, @ 60Hz >-25dB
Common mode rejection
0-60Hz -100dB
Input impedance
100 Mohm
±
10% (each differential input to ground)
Maximum tare cal offset
(Maximum tare offset depends on A/D range and SCP gain)
A/D range
±
V F.Scale
16
4
1
0.25
0.0625
Max Offset
.04970
.01220
.00297
.00055
n/a
10 HP
E1509
Fixed
Amplifier SCP
Specifications