TM 11-6625-2958-14&P
Table 5-8.
Checks and Adjustments After Replacement of Semiconductor Devices (Continued)
REFERANCE
I
FUNCTION OR CIRCUIT
CR90 thru
Crowbar.
CR93,
A4CR108,
A4CR110
A4CR101 thru
Main rectifier diodes.
A4CR104
A4CR105
Reverse voltage protection.
and
A4CR106
VR1
Voltage clamp circuit.
VR40
Mixer amplifier stabiliza-
tion diode.
CHECK
Limiting action and level.
Output voltage, ripple imbalance, and
preregulator waveforms.
Trip voltage, voltage across series regu-
lator when crowbar is tripped, supply
stability.
Voltage across main filter capacitors.
Output voltage.
CC load regulation.
CV transient response.
+6.2V and -6.2V reference voltages.
Trip voltage.
5-73 ADJUSTMENT AND CALIBRATION
5-74 Adjustment and calibration may be required
after performance testing, troubleshooting, or re-
pair and replacement.
Perform only those adjust-
ments that affect the operation of the faulty cir-
cuit and no others.
.5-76 The meter pointer must rest on the zero cal-
ibration mark on the meter scale when the instru-
ment is at normal operating temperature, resting
in its normal operating position, and turned off.
To zero set the meter proceed as follows:
a.
Connect load resistor of value shown in
b. Turn on instrument and allow it to come
up to normal operating temperature (about 30 min-
utes).
c.
Turn instrument off.
Wait one minute for
power supply capacitors to discharge completely.
d. Insert sharp pointed object (pen point or
awl) into small indentation near top of round black
plastic disc located directly below meter face.
e.
Rotate plastic disc clockwise until me-
ter reads zero, then rotate counterclockwise
ADJUST
- - -
R70, R82
R95,
A5R125
- - -
- - -
- - -
R47
- - -
R95,
A5R125
slightly in order to free adjustment screw from
meter suspension. Pointer should not move during
latter part of adjustment.
5-78 To calibrate the voltmeter, proceed as fol-
lows:
a.
Connect differential voltmeter across
supply, observing correct polarity.
b. Turn on supply and adjust VOLTAGE con-
trols until differential voltmeter reads exactly the
maximum rated output voltage.
c.
Adjust R106 until front panel voltmeter
also indicates exactly the maximum rated output
voltage.
5-80 To calibrate the ammeter, proceed as fol-
lows:
a.
Connect test setup shown in Figure 5-8.
b. Turn VOLTAGE controls fully clockwise.
c.
Turn on supply and adjust CURRENT con-
trols until differential voltmeter reads 0.5Vdc.
d. Adjust R101 until front panel ammeter in-
dicates exactly maximum rated output current.
Содержание 6260B
Страница 70: ...Figure 7 4 Bottom Front Chassis Assembly Component Location Diagram 7 4 TM 11 6625 2958 14 P ...
Страница 80: ...SECTION II MAINTENANCE ALLOCATION CHART TM 11 6625 2958 14 P D 3 ...
Страница 83: ...Figure 7 11 Schematic Diagram Model 6269B ...
Страница 88: ......
Страница 89: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ...THE METRIC SYSTEM AND EQUIVALENTS ...
Страница 95: ...PIN 046413 000 ...