3
Troubleshooting the MSD
General symptoms
69
Fan for the high vacuum pump is not operating
This refers to a condition where the MSD is on (at least the foreline pump is
operating) but the fan for the high vacuum pump is not working. You can deter-
mine this by listening or by putting your hand next to the ventilation slots on the
front of the instrument. You should be able to feel air being pulled into the MSD.
•
Disconnected fan cable
•
Disconnected high vacuum signal cable (
never
connect or disconnect
this cable while the MSD is on)
•
Malfunctioning fan
MSD is on but the status LEDs are all blinking
This is normal when the MSD is initially turned on. The status LEDs blink until the
data system software is started or the pumpdown cycle is initiated. The data sys-
tem then downloads the instrument control software to the HP-IB/MS control
card. When this occurs, the status LEDs will all light for a few seconds and then
begin to operate normally, leaving the power LED on and the others off. If the sta-
tus LEDs do not stop blinking, or if the status LEDs are all blinking at a time other
than startup, one of the following may be the cause:
•
Temporary power failure interrupted communications
•
Bad connection between the MSD and the data system
•
Software malfunction
•
MSD electronics are not working correctly
If the status LEDs continue to blink there will usually also be an error message in
the MSD ChemStation software.
MSD turns on but then the foreline pump shuts off
MSDs with serial number XX820XXXXX and above will shut down both the fore-
line pump and the high vacuum pump if the system fails to pump down correctly.
This is usually because of a large air leak: either the sideplate has not sealed cor-
rectly or the vent valve is still open. This feature helps prevent the foreline pump
from sucking air through the system, which can damage the analyzer and pump.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...