2
Operating the MSD
To tune the MSD
To tune the MSD
Software changes
The software is revised periodically. If the steps in this procedure do not match
your MS ChemStation software, refer to the manuals and online help supplied with
the software for more information.
1
In the Instrument Control View, select
Perform MS Autotune
from the
Instrument menu.
2
Select the tune program you wish to use.
The tune will start immediately. For most applications,
Autotune
gives the best
results.
Standard Spectra Tune
is not recommended, as it may reduce sensitivity.
Quick Tune
is used to adjust peak width, mass assignment, and abundance, without
changing ion ratios. Always tune the MSD with the same column flow and the
same analyzer temperatures that will be used for data acquisition.
3
Wait for the tune to complete and to generate the report.
Save your tune reports. To view history of tune results, select
View Tunes...
under
the Qualify menu.
4
To manually tune your MSD or to perform special autotunes, select
Manual Tune
from the View menu.
In the Manual Tune view, you can manually adjust most tune parameters to suit
special needs.
From the Tune menu, in addition to the tunes available from Instrument Control,
you can select special autotunes for specific spectral results:
DFTPP Tune
,
BFB Tune
, or
Target Tune
.
See the manuals or online help provided with your MSD ChemStation software for
additional information about tuning.
Содержание 5973
Страница 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Страница 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Страница 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Страница 30: ...30 ...
Страница 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Страница 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Страница 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Страница 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Страница 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Страница 92: ...92 ...
Страница 103: ...4 Maintaining the MSD To check and add foreline pump oil 103 Lower line Oil level window Fill cap Upper line ...
Страница 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Страница 107: ...4 Maintaining the MSD To refill the foreline pump 107 Fill cap Upper line Oil level window Drain plug ...
Страница 109: ...4 Maintaining the MSD To replace the oil trap 109 Oil trap Clamp O ring assembly Oil trap adapter Foreline pump ...
Страница 111: ...4 Maintaining the MSD To check the diffusion pump fluid 111 Fan Fluid level window sight glass ...
Страница 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Страница 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Страница 123: ...4 Maintaining the MSD To remove the calibration vial 123 Calibration valve Collar Calibration vial ...
Страница 125: ...4 Maintaining the MSD To refill and reinstall the calibration vial 125 Calibration valve Collar Calibration vial 6 mm ...
Страница 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Страница 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Страница 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Страница 210: ...210 ...
Страница 211: ...6 GC MSD Interface This chapter describes the function of the GC MSD interface GC MSD interface 212 ...
Страница 214: ...214 ...
Страница 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Страница 280: ...280 ...
Страница 294: ...294 Index ...
Страница 296: ...294 Chapter ...
Страница 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...