Chapter 4 Command Reference
Common Commands
Programming Guide
4-55
4
*SRE
(Service Request Enable Command)
*SRE?
(Service Request Enable Query)
COMMAND
*SRE <NRf> | <non-decimal numeric>
*SRE?
Sets or queries the Service Request Enable Register, which is shown in
Figure 4-5.
The parameter and query response value, when rounded to an integer
value and expressed in base 2 (binary), represents the bit values of the
Service Request Enable Register.
This register is used to enable a single or inclusive OR group of Status
Byte Register events to generate an SRQ.
Figure 4-5. The Service Request Enable Register
See the section titled “Status Byte Register and Service Request Enable
Register” starting on page 3-27 for a detailed description of the Service
Request Enable Register.
<NRF> RANGE
•
0 to 255
•
The value of bit 6 is ignored.
<NRF> RESOLUTION
1
QUERY RESPONSE
•
Numeric data transferred as ACSII bytes in <NR1> format.
•
The value of bit 6 is zero when queried.
Not Used
0 to 1
2
3
4
5
6
7
Bits
QSB
MAV
ESB
EAV
OSB
Not Used
Содержание 53150A Series
Страница 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Страница 2: ......
Страница 10: ...Contents x Programming Guide ...
Страница 11: ...1 Before You Start ...
Страница 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Страница 21: ...2 Command Summary A Quick Reference ...
Страница 41: ...3 Programming Your Counter for Remote Operation ...
Страница 100: ...Chapter 3 Programming Your Counter for Remote Operation Programming Examples 3 60 Programming Guide 3 ...
Страница 101: ...4 Command Reference A Dictionary ...
Страница 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Страница 163: ...5 Errors ...
Страница 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Страница 184: ......
Страница 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...