HP 456A Скачать руководство пользователя страница 17

Section  IV 
Paragraph 

4 - 1 0  

to  4- 1 3  

the  elimination  of  ground  loops .  If,  after  eliminating 

the  ground  loops ,  the  ripple  is  still  too  great  check 

CR92  and  Q90. 

Diode  CR92  should  have  14  volts 

volts  ac ross  it. 

8) 

if  the  instrument  section  is  faulty,  locate  the 

trouble  by  measuring  the  vol tages  as  shown  on  the 

schematic  diagram. 

CAUTION 

TAKE  CARE  NOT  TO  SHORT 

TH E 

WIRING 

WHEN  TROUBLESHOOTING.  EVEN  A  MO­
MENTARY  S HORT MAY BURN OUT A TRAN­

S ISTOR 

OR  mODE. 

The  transis tors  can  be checked  in  a  similar  manner. 
Since  the  transistors  used  in  this  instrument  are of 

the  pnp  type  the  voltage  at  the  emitter  should  be 
slightly  positive  (w ithin  0.4  volt) of the voltage on  the 

base.  The  vol tage  on  the  collector  should  be  nega­
tive  with  respect  to  the  base  and of the value  shown 

on the  schematic  diagram. 

Unfortunately ,   due  to  the  large  amount  of  feedback 

any  trouble  may  cause  all  of  the  voltages  to  be off. 
First,  try  to find  the  trouble  by  resistance  measure­

ments . 

If  this  fails  you  may  have  to  remove  the 

transistors  and  diode,  and  test  them  on  a  simple  dc 
tester . 

4 -1 0 .  

S EMI - C O N DU CTOR  RE P LACEME N T .  

Do 

NOT  replace  the  transistors  or  diodes  unless 

there  is  a  definite  indication  that  these  components 
are  faulty,  as  you  may  damage  them.  Turn  the  in­
strument  OF F  before  doing  any  soldering  as  ac 
leakage  from  the  soldering  iron  may  overload  and 

damage  the  transistors .  In  addition  before using an 

iron  on  any  semi-conductor  lead,  clip a  pai r  of long­
nosed  pliers  between  the  iron  and  semi-conductor . 

The  pair  of  pliers  will  act as  a  heat  sink and  absorb 

hear  helping  to  prevent  damage  to 

the 

semi- conductor . 

Table  4-1  lists  the  tes ts  which  should  be  made after 

replacing the components . 

4 -11 . 

CAL I BR AT I O N . 

Perform  the  procedure  in  figure 

4-2 

at  1  kc .  Adjust 

R9,  Gain  Set  Potentiometer,  so  that  the  voltage  out 
is  exactly 

0. 1 

volt  (millivolts  output  equals  milli­

amperes  input). 

If 

you can't  reach  calibration,  check 

R7 , 

9,  1 0  and  1 2,  and  the  probe . 

4 - 4  

Model  456A 

4 -12. 

F R EQUEN C Y   R ESP O N S E. 

Perform  the  procedure  given  in figure  4- 2. 

4 -1 3. 

R E P LAC I N G   T H E   PRO B E   C A B L E .  

A .  GENERAL. 

The  cable  assembly  used  i n   the 

456A-2 1 A   P robe  Assembly  includes  in  the  probe 

cable  assembly  a  speCial  cable  terminated  at  one  end 
with  a  Mic rodot  miniature  connector and  at  the  other 

end  with  a  crimp- on ground lead .  Both  of  these con­
nectors  may  be  replaced  but it is  recommended  that 

a  new  probe  cable  as sembly, 

� 

s tock # 456A - 2 1 A - 8, 

be installed instead  because of difficulties encountered 
in  replacement. 

B .  

MIDRODOT  M INIATURE  CON N ECTOR. 

The 

microdot  connector  terminates  the  end  of  the 

cable  under  the  handle  (see  figure 4- 3).  To  replace 
this  connector  four  new  parts  and a  special hand tool 
for  ass em bl y   a re  required.  See  the  operating  note 
for  the 

c!j!l 

425A- 2 1 B   Microvolt  Ammeter  P robe  for 

ins tructi ons . 

C.  CRIM P -ON  GROUND  CON NEC TOR.  This  con-

nector  is  crimped  to avoid  soldering and conse­

quent  melting of the cable ins ulation .  Crimp the ground 
lead  on  again.  if  possible.  If  the  ground  lead  must be 

soldered,  clamp a  pair  of pliers  between  the  iron  and 
insulation  to  act  as  a  heat  sink. 

Table 

4- 1 .  

Replacement 

Componen 

Test 

CRI 

None 

CR90 

Check power  supply  (pa r .   4-

9) 

CR91 

Check power  supply  (pa r.  4-9) 

CR92 

Check power s upply  (pa r.  4 - 9) 

Q l  

None 

Q2 

Check  calibration  (par . 

4- 1 1 )  

Q9

Check  power s upply  (par.  4-9) 

P robe 

Check calibration  (par.  4- 1 1); 

check frequency  response  (par . 

4 - 1 2: 

00395- 1 

Содержание 456A

Страница 1: ...56A AC CURRENT PROBE SERIALS PREFIXED 103 PRINT O 3 61 I 00395 1 ...

Страница 2: ...00395 1 OPERATING AND SERVICING MANUAL MODEL 456A SER IALS PREFIXED 103 AC CURRENT PROBE Copy ht HEWLETf ACK UD COM ANY 1961 ISOI AGE Mill ROAD AlO AlIO CALIFORNIA U S A Printed MAR 1961 ...

Страница 3: ... 13 Replacing the Probe Cable REPLACEABLE PARTS 5 1 5 2 5 3 Introduction Table of Replaceable Parts Ordering Information 4 L 4 1 4 L 4 1 4 1 4 3 4 3 4 3 4 3 4 4 4 4 4 4 4 4 5 1 5 L 5 1 5 1 LIST O F IllUSTRATIONS Number Title L l General A rrangement of Model 456A 1 2 Model 456A AC Current Probe 1 3 Typical Change in F requency Response 2 1 2 2 2 3 2 4 2 5 with Load Capacitance Probe used with Mcxi...

Страница 4: ... probe Less than 50 milliohms in series with 0 05 h This is approximately the induc tance of 1 1 2 inches of hookup wire Approximately 4 pf added from wire to ground For 1 2 amp input at least 50 db down For 10 rna input at least 70 db down Less than 50 a rms 100 a when ac powered 220 ohms at 1 kc Approximately 1 volt dc component Should work into load of not less than 100 000 ohms shunted by appr...

Страница 5: ...or The probe operates with an accompanying small ampli fier to convert the ac current being measured to a pro portional voltage The current to voltage conversion factor is 1 millivolt output for 1 milliampere flowing th rough the probe Current readings can thus be taken directly from the voltage calibrations on the voltmeter or oscilloscope Table 1 2 is a list of instruments w hich can be used wit...

Страница 6: ... The Model 456A is available either as a battery powered or ac powered instrument The AC Power Supply srock 456A 95B fits into the space nor mally occupied by the battery supply and takes the place of it electrically The ac power supply must be ordered separately When the 456A is powered with its own internal bat teries it is independent of power lines It may be used in the field with battery oper...

Страница 7: ...ribed in text 00395 1 In this manner currents even below 10 microamperes can be measured 2 3 OPERAT IN G T HE PRO BE The probe jaws may be operated with one hand While holding the probe handle in the palm squeeze the flanges together with the fingers This action will open the jaws Move the probe to the vicinity of the insulated wire being measu red and slowly release the flanges until the jaws fit...

Страница 8: ...by squeezing flanges on probe together 3 Connect ground clip to instrument under test if necessary see paragraph 2 5E Figure 2 2 2 2 Model 456A 4 t 1I t f RO 4 Read current being measured on meter or oscil loscope read millivolts as milliamperes S Note direction of arrow on probe Conventional current flow in direction of arrow gives positive going output voltage Opera tion 00395 1 ...

Страница 9: ...n the signal being measured should preferably be kept below 1 2ampere Direct current below that value will have no effect on the measurement C ALTERNATING CURRENT FIELDS The probe is shielded against external ac magnetic and electric fields Extremely strong fields may cause an erroneous reading If a strong ac magnetic field 00395 1 is suspected rotate the empty probe head with the jaws closed If t...

Страница 10: ...u may choose which side of the circuit to measure the current See figure 2 5 Note in figure 2 5 the probe measures the current flowing in the circuit on the same side as the wide side of the arrow on the probe Occasionally at high frequencies above 15 me a greatly magnified voltage effect exists at a particular frequency This effect is caused by a resonance be tween excessive external ground lead ...

Страница 11: ... to the tube output capacity hp 456A PROBE HEAD Section II Paragraph 2 6 Cont d The probe is valuable in a number of other current equalizing applications EspeciaUy common among these are equalizing the input and output currents in push pull and balanced circuits in both transistor and tube applications Figure 2 6 Bucking OUt Capacitive Component of Plate Current 00395 1 2 5 2 6 ...

Страница 12: ...results in a better frequency response lower input impedance dc bias stability lower distortion etc le 4 RD Figure 3 1 Equivalent Circuit 00395 1 The equivalent circuit can be thought of as shown in figure 3 1 where CL is the current gain of the tran sistor The transistors have internal resistances as shown Emitter resistance re is small but relevant to this discussion while the collector resistan...

Страница 13: ...lowed extensive damage to the etched circuit board will result 1 Apply heat sparingly to lead of component to be replaced If lead of component passes through an eyelet in the circuit board apply heat on com ponent side of board If lead of component does not pass through an eyelet apply heat to con dUctor side of board 3 Bend clean tinned leads on new part and care fully insert through eyelets or h...

Страница 14: ...tion unless otherwise specified It can be quickly and easily converted to operate from a nominal line 00395 1 voltage of either 115 or 230 volts and a frequency of 50 to 1000 cps To convert for 230 volt operation remove the short across R90 Turn the instrument off before doing any soldering 4 4 C HECKIN G THE BATTERIES Whenever trouble is encountered check the voltage at the BATTERY TEST terminals...

Страница 15: ...il400D H L reads exactly 0 1 volt Model 456A 456A 5 Connect the OUTPUT cable of the 456A to the input of the 400D H L in place of the 470C Leave the wires still connected to the 470C 6 Read 400D H L Reading should be within sum of specifications of the 470C and 456A 7 Repeat above steps with 650A set at other fre quencies up to 100 kc If the 456A calibration or frequency response is poorer than te...

Страница 16: ...illoscope should be less than 75 Jl volts 100 Jl volts with ac powered model If not check C94 power supply capacitor if ac powered or try replacing CRI Q1 or Q2 Check with 456A POWER switch off to see if ground loops exist If they do see paragraph 2 50 B CALIBRATION A ND FREQUENCY RESPONSE Perform the procedure illustrated in figure 4 2 Note that this test is used only for rough testing with readi...

Страница 17: ...o the semi conductor Table 4 1 lists the tests which should be made after replacing the components 4 11 CALIBRATION Perform the procedure in figure 4 2 at 1 kc Adjust R9 Gain Set Potentiometer so that the voltage out is exactly 0 1 volt millivolts output equals milli amperes input If you can t reach calibration check R7 9 10 and 12 and the probe 4 4 Model 456A 4 12 FREQUEN CY R ESPO N S E Perform ...

Страница 18: ...TEXT LOCKWASH ER l 2190 0012 PROBE CABLE ASSEMBLY COMPLETELY ASSE MBLED 4 5 6A 2 IA B PROBE G R OU N D LEAD ASSEMBLY CONTAI NS 2 2 OHM RESI STOR 456 A 2I A 10 PROBE F L CENTER SECTION COIL SP R I NG TO C3 i FRONT PANEL I I O F I NSTRUMENT LI INTERNAL LOCKWASH E R Figure 4 3 Exploded View of P robe Assembly Section IV Figure 4 3 00395 1 4 5 ...

Страница 19: ...T 1101 OPEN leOV K T9 II o 11 0 Y 111 0 f nUIA t1191 r Y IV WK 011 1110 wlff v NOMINAL 11 0 2 J TO ull o RII 2 7 BL f WHT A 860 C90 4OI1F O OI f e5 O OI F e6 10UF R7 3300 AC POWER SUPPLY Rn 90 10K 40 IIf R93 6100 c 1 Q Q90 2N650 REGUL ATOR 131 BATTERY SUPPLY IIC IJLATC D r o 811 lIT 1 T lIT 2 f oOV BTl C O S 4V n 14 R IZ 100 U1 Figure 4 4 Model 456A AC Current Probe R9 ICO e93 lOll RIO 100 RI lao ...

Страница 20: ......

Страница 21: ... 3 O R DERI N G IN FORMAT I O N To order a replacement pa rt address order or in quiry either to your authorized sales office or to CUSTOMER SERVICE Hewlett Packard Company 395 Page Mill Road Palo Alto California or in Western Europe to Hewlett Packard S A Rue d u Vieux Billa rd No 1 Geneva Switzerland Specify the following information on a part 1 Model and serial number of the instrument Be s u r...

Страница 22: ...Diode germanium IN38A Diode silicon Connector female red Connector female black Connector male output Not assigned Cable power Transistor 2N1 516 0C1 70 ot assigned rransistor 2N6 50 esistor fixed composition 24 ohms 5 1 W esistor fixed composition 22 ohms Part of Probe Ground Lead Assembly Component not separately replaceable Mfr 1 Stock No 9 609 5 01 50 0008 5 6289 0 180 0050 5 62 89 0180 0061 5...

Страница 23: ...igned Resistor fixed composition 1 5 000 ohms 10 2 W Resistor fixed composition 680 oluns 10 1 2 W Resistor fixed composition 10 000 ohms 1 0 1 2 W Resistor fixed composition 6800 oluns 10 1 2 W Resistor fixed deposited carbon 4000 ohms 1 1 2 W Resistor fixed depOSited carbon 2 2 50 0luns 1 1 2 W Switch slide DPDT Not assigned Transformer powe r Section V Replaceable Parts Sheet 3 of 4 Mfr J Stock...

Страница 24: ...onnec tor male Can shield Boot cable Spacer standoff Lead probe ground includ e s R 5 Replaceable Parts Sheet 4 o f 4 Mfr 2 8480 2 8480 2 8480 i 99109 24655 24655 28480 28480 00866 28480 I F Stock No G 64B G 64C 8520 0017 456A 16A 8 120 0048 1 2 50 009 1 1 250 0092 4 56A 55A 4 12A 8 3A 0380 0048 4 56A 2 IA J rQ R S 1 0 1 0 1 0 1 1 1 0 1 0 1 0 1 0 1 0 1 0 1 1 See introduction to this section Model ...

Отзывы: