SECTION II   

OPERATING INSTRUCTIONS 

 
2-1

LOW-LEVEL ELECTRICAL PHENOMENA.   

2-2. Stray low-level electrical phenomena are present in one form or another, in nearly all electrical circuits.   
The 412A does not distinguish between stray and signal volta

g

es

í

measures net voltage. Thus, when using   

the lower voltage ranges, consider the pos sib ilit 

of low -level electrical phenomena. Thermocouples   

(thermoelectric effect), flexing of coaxial cables (triboelec

t

ric effect), apparent residual charges on capacitors   

(dielectric absorption), battery action of two terminals mounted on an imperfect insulator (galvanic action)   
all can produce voltages within the range of the 412A

.   

 

2-3. The 412A voltage probe, current/resistance lead and common lead are designed to have a very low thermoelectric 
effectwith copper, the most complete electrical conductor. However, you may encounter other materials. For example,   
the leads of many transistors are made of a rníxture of iron, nickel and cobalt known commercially as Kovar, Fernico,   
etc. This material makes a very good thermocouple with copper: about 40 

J.l 

/

oC with respect to a reference junction.   

 
2-4. Whenever possible, connect the 412A leads to copper and maintain the points of connection at the same   
temperature, preferably ambient temperature. With the leads so connected, any voltage indicated by the 412A is   
developed within the circuit under test.   
 
2-5. 

OPERATING INSTRUCTlONS.   

 

 

DO NOT OVERT~OAD THE INSTRUMENT. AMPLIFIER INPUT, CURRENT SHUNTS,   
AND INTERNAL RESISTANCE STANDARDS ARE NOT PROTECTED FROM EXTREME   
OVERLOAD. MOMENTARY OVERLOADS TEN TIMES FULL SCALE WILL NOT   
DAMAGE THE INSTRUMENT.   
 
2 -6. Turn the Model 412A on and allowa few minutes warmup.   
 
2-7. VOLTAGE MEASUREMENT.   

a.

 

Set FUNCTION selector to VOLTS.   

b.

 

Set POLARITY switch to desired polarity.   

c.

 

Set RANGE switch to desired range.   

d.

 

Use VOLTS and COM leads to connect ínstru-   
ment across circuit or component, and read   
voltage.   

2-8. CURRENT MEASUREMENT.   
a. Set FUNC TION selector to MA.   
b

Set RANGE switch to desired range.   

c. De-energize circuit to be tested.   
d. Use MA/OHMS and COM leads to connect instrument into circuit.   
e. Energize circuit, set POLARITY switch for upscale reading, and read current.   

 

--------------NOTE-------

-

----- 

When measuring current, be sure there is no connection between the chassis-ground   
and cabinet-ground terminals of the DC AMPLIFIER OUTPUT connector.   
 

2-9. RESISTANCE MEASUREMENT.   

a. Set FUNCTION selector to OHMS.   
b. De -energize circuit to be measured.   

c.

 

Use MA

/

OHMS and COM leads to connect instrument across circuit or component.   

d.

 

Select range which brings meter pointer as close as possible to midscale, and read resistance.   
 

2 -10. When measuring the resistance of non-linear devices such as crystal diodes or transistors, you   
may want to know the voltage applied to the device and/or the current through it at the time of measurement   
By using Table 2 -1 and the upper voltage scale on the meter face, you can determine both. The meter   
reading is directly proportional to the voltage across the device being measured, and the meter reading   
subtracted from full scale is directly proportional to the current. For example, on the X10 range the supper   
voltage scale is a 0-100 mv scale and a 10-0 ma scale. If the meter indicates .25 on the resistance scale

the   

device being measured has an equivalent resistance of 2.5 ohms. But the meter also indicates .2 on the   
upper voltage scale; thus, from Table 2 -1, the device has 2. 5-ohms equivalent resistance with 20 mvacross   
it and 8 ma flowing through it.   
 
2-11

OPERATION WITH A RECORDER.   

 
 
 
 
 

Содержание 412A

Страница 1: ...sociation The third contact is an offset round pin added to a standard two blade connector This contact grounds the instrument when used with an appropriate receptacle An adapter should be used to connect the NEMA plug to a standard two contact output When the adapter is used the ground connection be comes a short lead from the adapter This lead should be connected to a suitable ground for the pro...

Страница 2: ...allowa few minutes warmup 2 7 VOLTAGE MEASUREMENT a Set FUNCTION selector to VOLTS b Set POLARITY switch to desired polarity c Set RANGE switch to desired range d Use VOLTS and COM leads to connect ínstru ment across circuit or component and read voltage 2 8 CURRENT MEASUREMENT a Set FUNC TION selector to MA b Set RANGE switch to desired range c De energize circuit to be tested d Use MA OHMS and C...

Страница 3: ...es modulator output about 500 000 times A demodulator synchronized with the modulator by the light beam chopper converts amplifier output to de The output of the demodulator is filtered and applied through a cathode follower to 1 a feedback network 2 the DC AMPLIFIER OUTPUT terminals and 3 an output indicator The feedback network stabilizes the dc gain of the modulator amplifierdemodulator system ...

Страница 4: ...2 Reheat solder in vacant eyelet and quickly insert a small awl to clean inside of hole If hole does not have an eyelet insert awl or a 57 drill from conductor side of board 3 Bend clean tinned leads on new part and carefully insert through eyelets or holes in board 4 Hold part against board avoid overheating and solder leads Apply heat to component leads on correct side of board as explained in s...

Отзывы: