Section V
,
.
Model
403A
Paragraphs
5-7 to 5-8
Figure
5-2.
Model
403A
Bottom
View
5-7.
ACCURACY
T E S T.
d. Check the other ranges using the voltmeter
c a l i b r a t o r
outputs andModel 403A indications
5-8. A voltmeter calibrator
(-hp-
Model
738BR
Vo l t -
listed
in
Table 5-2.
meter
Calibrator) is r e q u i r e d f o r the accuracy
test.
a.
Connect
voltmeter
calibrator outputto Model
Table
5-2. Accuracy
Te s t
403A
input
a s
shown
i n F i g u r e
5-3.
VOLTMETER CALIBRATOR
Voltmeter
C a l i b r a t o r
Model
403A
hp738BR
hp
403A
Output
Indication
VOLTMETER
(400
cps
r m s )
(4 3% of
f u l l
scale)
Figure
5-3.
Accuracy
Te s t
b.
Set
Model
403A FUNCTION to 15
C P S - 1
M C .
c.
Set
voltmeter calibrator output to
300
V,
400
cps
r m s . Model
403A should indicate
300
V
+3%
(Table 5-2).
5-2
00125-4