Section V
calibrated, set the REL. REF
adjustment to the fully clockwise
ABSOL U T E posi tion before
beginning the calibration. Always
check and set the bias levels first
before calibration.
5-41. Accuracy and Frequ ency Respons e Calibration.
5-10
a. Calibrate the reference supply in Figure 5-2
with a 1 volt thermal converter according to
the steps in Paragraph 5-17.
b. Disconnect the standard and connect the
test oscillator, the 400E/EL, and the digital
voltmeter as shown in Figure 5-2. Connect a
precision 40dB coaxial attenuator between
the BNC "T" connector and the 400E/EL
INPUT. When the thermal converter and
reference supply outputs are nulled, the
input to the 400E/EL will be 10mV. Set the
oscillator frequency to 400Hz and the
400 E/ EL to the 0.01 volt range. Using the
amplitude control as coarse adjustment and
R4 as fine adjustment, increase the oscillator
output until the thermal converter output
nulls the reference supply.
c. Adjust A2R38 for a 400E/EL meter reading
of IOmV ±O. l mV or less.
d. Ad just R2 for a digital voltmeter display of
1.000 ±0.005Vdc or less.
e. Lower test oscillator output and set
frequency to IOMHz. Readjust oscillator
amplitude until thermal converter output
nulls reference supply.
f. Adjust A2C28 for digital voltmeter display
of 1 .000 ±0.04 or less.
g. Lower test oscillator output and set the fre
quency to 10 MHz. Disconnect the precision
40 dB attenuator and connect a precision
50 dB attenuator in its place. When the
thermal converter and the reference supply
are nulled, the input to the 400E/EL will be
3.162
mY.
Set the 400E/EL RANGE to
0.003 volts.
h. Using the amplitude control as a coarse
ad justment and R4 (Figure 5-2) as a fine
adjustment, increase the oscillator output
until the thermal converter exactly nulls the
reference supply output.
Models 400E/400EL
i.
Adjust S2C2 for a digital voltmeter
indication of 1.000 ±O.04Vdc. The 400E/EL
should read 3.16 ±O. l 58mV.
Check at 6 and 4MHz. Amplitude
can be lowered at 4MHz by mOving
A2Cl i and C13 closer together.
j. Lower the test oscillator output and set the
frequency to 400Hz. Replace the 1 volt
thermal converter with a 0.45 volt thermal
converter, and calibrate it according to the
procedure in Paragraph 5-17 using a 0.3162
volt output from the dc standard.
k. Disconnect the dc standard and connect the
test oscillator to the thermal converter as
shown in Figure 5-2. Set the 400E/EL range
switch to 0.001 volts. Using the amplitude
control as a coarse adjustment and R4 as a
fine adjustment, increase the oscillator
output until the thermal converter output
exactly nulls the reference supply output. At
null, the input to the 400E/EL will be 1 mV
at 400Hz.
m. Ad just A2R44 for a digital voltmeter display
of 1 .000 +0.005 V dc.
The voltage at TP 4 affects fre
quency response. If the frequency
response is low at 10 and 20 Hz
decrease voltage at TP4 (extreme
limits 8 to 12 volts). After changing
voltage at TP4, check high and low
frequency response.
n. If the accuracy and frequency response can
not be brought into specifications with the
preceeding adjustments, refer to Table 5-11
Factory Selected Components.
5·42. Calibration Procedure Using -hp- 745A1746A
Calibrator.
5-43. Calibrate the 400E/EL for frequencies from
10Hz to 110kHz with the test setup shown in Figure
5-1.
a. Set the 745A to the voltage levels and
frequencies listed in Table 5-5, Alternate
Calibration Procedure (omit steps 2 and 3).
Select the proper 400E/EL range.