Hovertech Hoverjack Evacuation Скачать руководство пользователя страница 2

User Manual

2

  |  HoverTech International  

EVHJ Manual Rev. B

Intended Use and Precautions

INTENDED USE

The Evacuation HoverJack® Device is used to transport patients up or  

down stairs in the event of an emergency. The Evacuation HoverJack can 

also be used to lift a patient in supine position from the floor to bed or 

stretcher height, utilizing the HoverTech Air Supply to inflate each of the 

four chambers.

INDICATIONS

•  Patients unable to assist in their own vertical lift or evacuation, such as 

after a fall or in an emergency situation

•  Patients whose weight or girth poses a potential health risk for the 

caregivers responsible for lifting or moving said patients

CONTRAINDICATIONS

•  Patients who are experiencing thoracic, cervical or lumbar fractures 

that are deemed unstable, unless using in conjunction with a spinal 

board (follow your state’s protocol regarding use of spinal boards)

INTENDED CARE SETTINGS

•  Hospitals, long term or extended care facilities

PRECAUTIONS

•  Make sure patient safety straps are secured before moving.
•  Move the Evacuation HoverJack using the transport straps and/or the 

transport  handles located along the top perimeter.

•  Never use patient safety straps to pull the Evacuation HoverJack, as 

they may tear.

•  When moving a patient on the inflated Evacuation HoverJack, use 

caution and move slowly. 

•  Additional caregivers are recommended when moving or evacuating a 

patient over 250 lbs./113 kg. 

•  Never attempt to move a patient on an uninflated Evacuation  

HoverJack.

•  Never leave a patient unattended on an inflated device.
•  Use this product only for its intended purpose as described in this 

manual.

•  Only use attachments and/or accessories that are authorized by  

HoverTech International.

WARNING/CAUTION

•  For safety, always use a minimum of two caregivers while using the 

Evacuation HoverJack.

•  Reference product-specific user manuals for additional operating 

instructions.

PRECAUTIONS   

– HOVERTECH INTERNATIONAL AIR SUPPLY

•  Not for use in the presence of flammable anesthetics or in a hyperbaric 

chamber or oxygen tent.

•  Route the power cord in a manner to ensure freedom from hazard. 
•  Avoid blocking the air intakes of the HoverTech International Air 

Supply.

•  CAUTION: Avoid electric shock. Do not open HoverTech International 

Air Supply.

Symbol Reference

DECLARATION OF CONFORMITY  

TO MEDICAL DEVICE DIRECTIVE

Содержание Hoverjack Evacuation

Страница 1: ...n Evacuation HoverJack Device 3 Part Identification HT Air Air Supply 4 HT Air Keypad Functions 4 Air200G Air400G Air Supplies 4 Instructions for Use 5 Product Specifications Required Accessories 6 Cl...

Страница 2: ...k using the transport straps and or the transport handles located along the top perimeter Never use patient safety straps to pull the Evacuation HoverJack as they may tear When moving a patient on the...

Страница 3: ...dend Transport Strap 2 Inflatable Wedge with Inlet Cap and Valve Red Deflate Cap and Valve Patient Safety Straps 4 DO NOT USE TO TRANSFER Accessory Pouch Label includes Model Serial and Instructions f...

Страница 4: ...l indicate the inflation speed by the number of flashes i e two flashes equals the second inflation speed All of the settings in the ADJUSTABLE range are substantially lower than the HoverMatt and Hov...

Страница 5: ...uation HoverJack WARNING DO NOT RELEASE MULTIPLE CHAMBERS AT ONCE Evacuation HoverJack Device Instructions for Use as an Evacuation Device 1 To inflate follow steps 1 4 of instructions for use as air...

Страница 6: ...p The red caps are intact All patient safety straps are attached All buckles and zippers if applicable are intact and operational Transport handles and straps are all attached Inlet valves are all sel...

Страница 7: ...m of any kind for any loss or damage arising out of connected with or resulting from the design manufacture sale delivery installation repair or operation of its products whether in contract or tort i...

Страница 8: ...User Manual 8 HoverTech International EVHJ Manual Rev B 4482 Innovation Way Allentown PA 18109 800 471 2776 Fax 610 694 9601 www HoverMatt com Info HoverMatt com...