6.      How to use 

1)

 

Connect a gas line in the arrow direction given on the main body. 

2)

 

Make electrical connection in accordance with the Connector Pin Connections List. For 
the DC power, use the following capacity: 
+15V , 50mA and –15V , 200mA or more 

3)

 

After turning on the power, stop the gas and warm up the apparatus for about 5 minutes. 

4)

 

Start a gas supply a set voltage. Flow rate control will start according to the set voltage. 

5)

 

The set voltage to obtain a required flow rate value can be calculated by linear 
conversion. 

Required flow rate 

Set voltage    =                                                                     

×

 5.000V 

                                                                        Full-scale flow rate 

6)

 

Forced full opening(purging) / closing of the SEC-E400X main body control valve: The 
valve is fully opened by connecting the supply voltage of +15V to the No.1 pin(valve 
opening / closing input) of the main body connector, and fully closed by connecting the 
supply voltage of –15V. 

In the normal state, open the No.1 pin(nothing connected). 

    Otherwise, flow rate control may fail if a certain signal is input. 

7)

 

Full closing function 
The control valve of the SEC-E400X is fully closed by setting the voltage to 0(50mV or 
less). Control is resumed by applying the set voltage of 200mV or more. 

An input from the No.1 pin has the highest priority. 

8)    Valve Test Point ( Valve Voltage Monitor) 

The voltage applied to the control valve of the mass flow controller is output to between 
the No.8 and No.5 pins of the connectors. Note that the minus(-) side of the valve voltage 

        monitor has the potential of –15V with respect to the power COMMON terminal. 
                An output voltage range is 0V(fully closed) to +15V(fully opened). 

If the No.8 pin is monitored, referencing the power COMMON terminal, the difference 
b15V and the valve applied voltage will be output in terms of negative polarity. 
An output voltage range is –15V(fully closed) to 0V(fully opened).   
Since this voltage differs depending on the pressure or temperature used, monitor it as 
just a guideline. A monitor output is not linear to a flow rate since the control valve has 
hysterisis characteristics. 

                The voltage does not show an absolute amount. Assume that 0 to 15V correspond to 0 to 

100% of the voltage applied to the control valve. 

 
    Note: Description 4) to 8) are not applied to SEF-E431X/E441X. 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Valve Test Point 

15V 

Power COMMON

Control Valve 

Valve Voltage Monitor 

(+) 

(-) 

-15V Valve   
Applied Voltage 

Содержание SEC-E431X

Страница 1: ...MASS FLOW CONTROLLER MASS FLOW METER SEC SEF E431X E441X PTFE TYPE Instruction Manual CODE I031150800...

Страница 2: ...Jun 2012 2007 2012 HORIBA STEC Co Ltd...

Страница 3: ...CATIONS 3 PRINCIPLE OF FUNCTIONING 4 EXTERNAL DIMENSIONS 5 CONNECTOR 6 HOW TO USE 7 CONNECTIONS 8 PRECAUTIONS FOR HANDLING 9 PRODUCT WARRANTY 10 CHEMICAL CONTENT INFORMATION THE SIX CHINA RoHS RESTRIC...

Страница 4: ...0 kPa G Pressure resistance 6 1000 kPa G Leak integrity 8 1 10 9 Pa m3 sec He or less Setting signal 0 25 5V DC Input impedance 1 M or more Output signal 0 5V DC Minimum load resistance 2k Power sourc...

Страница 5: ...uracy linearity and repeatability are guaranteed for the calibration gas 4 Temperature for accuracy guarantee is in accordance with SEMI standard E56 1296 5 D indicates differential pressure G indicat...

Страница 6: ...4 4 External Dimensions SEC SEF E431X E441X Top Connector Type Side Connector Type...

Страница 7: ...est point 1 3 9 N C 4 1 N C no connection for SEF E431X E441X 2 In order to prevent COMMON voltage fluctuations due to a valve drive current the power COMMON terminal Pin No 4 and signal COMMON termin...

Страница 8: ...by setting the voltage to 0 50mV or less Control is resumed by applying the set voltage of 200mV or more An input from the No 1 pin has the highest priority 8 Valve Test Point Valve Voltage Monitor Th...

Страница 9: ...ired A control valve driving current max 200mA flows to the power COMMON line In order to prevent a voltage drop due to wiring be sure to wire the power COMMON line and signal COMMON line separately a...

Страница 10: ...trouble 8 Do not apply an excessive force to the main body or cable 9 Do not apply an excessive pressure to the main body 10 The withstand voltage for the set input is 15V Never apply a voltage beyond...

Страница 11: ...from the irresistible forces such as natural disasters and so on Troubles resulting from wrong handling When the apparatus is used or stored in an inadequate environment When the apparatus is used be...

Страница 12: ...ine parts SJ T11363 2006 Denotes that the amount of the toxic or hazardous substance contained in all of the homogeneous materials used in the component is below the limit on the acceptable amount sti...

Отзывы: