Data Management
63
Execution confirmation for [Save Calibration History]
This is a confirmation message for saving the calibration history to a USB memory stick.
Fig. 63 Execution confirmation for [Save Calibration History]
The buttons and button functions, which can be used with the execution confirmation for [Save
Calibration History] appearing, are described in the table below.
Table 14 Button functions with execution confirmation
for [Save Calibration History]
If the ENT button is pressed, the calibration history is saved to a USB memory stick as a file in
CSV format.
Each line describes one set of data. The values are separated by commas (",").
Saved items and formats are as shown in the table below.
Output example:
Date,Standard,Unit,Status
2001/01/01 12:34:56,0.0,mg/L,0
2001/01/02 12:34:56,200,mg/L,2
Button
Function
Page
ENT button
Saves the calibration history to a USB memory stick.
-
ESC button
Closes the message.
-
Item (1st line is the title)
Format
Remarks
Date
yyyy/mm/dd hh:MM:ss
Date and time of calibration (year, month, day, hour, minute,
second)
Standard
ddddd
Calibration concentration value (0 for zero calibration, set span
value for span calibration)
Unit
uuuuu
Units of calibration concentration (fixed at mg/L)
Status
sssss
Error flag
0: No error 2: Warming up