Troubleshooting
106
Troubleshooting
Alarm displays and actions
If an alarm occurs, the alarm icon at the top of the LCD flashes.
You can check the current alarm on the Current Alarm screen.
Reference
For information on the Current Alarm screen, refer to "Current Alarm screen" (page 58).
List of alarms
Display
Description
Cause
Action
System Failure
An error occurred due to
insufficient RTOS
resources.
An internal problem
other than those
below occurred.
Contact your local dealer or service station.
Clock
The clock was reset
because power was not
supplied to RTC during
startup.
The clock battery is
dead.
The date and time are initialized to 2001/01/
01 00: 00:00.
Refer to "System Setting screen" (page 84) to
set the date and time. The alarm will be
cleared until the power is turned OFF.
Contact your local dealer or service station to
replace the clock battery.
Load (Fact)
Failed to read factory
settings.
The internal memory
has failed.
Contact your local dealer or service station.
Load (User)
Failed to read general
settings.
The internal memory
has failed.
Contact your local dealer or service station.
Load (Meas)
Failed to read the
measurement history.
The internal memory
has failed.
Contact your local dealer or service station.
Load (Calib)
Failed to read the
calibration history.
The internal memory
has failed.
Contact your local dealer or service station.
Load (Alarm)
Failed to read Alarm
History.
The internal memory
has failed.
Contact your local dealer or service station.
Save (Fact)
Failed to save or delete
factory settings.
The internal memory
has failed.
Contact your local dealer or service station.
Save (User)
Failed to save or delete
general settings.
The internal memory
has failed.
Contact your local dealer or service station.
Save (Meas)
Failed to save or delete
measured values.
The internal memory
has failed.
Contact your local dealer or service station.
Save (Calib)
Failed to save or delete
the calibration history.
The internal memory
has failed.
Contact your local dealer or service station.
Save (Alarm)
Failed to save or delete
Alarm History.
The internal memory
has failed.
Contact your local dealer or service station.
Motor (U)
Hardware error in flow
path switching motor at
top of the main unit.
The motor has
failed.
Contact your local dealer or service station.
Motor (L)
Hardware error in flow
path switching motor at
bottom of the main unit.
The motor has
failed.
Contact your local dealer or service station.