Setting
84
System Setting screen
If [System Setting] is selected on the Setting Top screen, the System Setting screen opens.
The System Setting screen shows the current system settings.
If an item is selected with the up/down button and the ENT button is pressed, a pop-up screen
appears to let you change the setting of the selected item.
Fig. 94 System Setting screen
The setting items on the System Setting screen are as shown in the table below.
Table 27 Items on the System Setting screen
The buttons and button functions, which can be used with the System Setting screen
appearing, are described in the table below.
Table 28 Button functions with the System Setting screen
Note
The screen does not change, if [Ver.] is selected and the ENT button is pressed.
Item
Description
Setting range or selections
Page
Language
Sets the language.
ENGLISH, JAPANESE, RUSSIAN
page 85
B-Light Off Time
Sets the time of turning OFF the light of the
LCD and extraction tank automatically.
OFF, 10 min, 30 min, 60 min
page 85
Date
Sets the date.
2000/01/01 to 2099/12/31
page 86
Time
Sets the current time.
00:00 to 23:59
page 86
Ver.
Shows the software version (fixed value) in
the right column.
-
-
Button
Function
Page
ENT button
Shows the screen for setting the selected item.
-
Up button
Selects the next item up.
-
Down button
Selects the next item down.
-
ESC button
Returns to the Setting Top screen
page 69
Содержание OCMA-500
Страница 1: ...Oil Content Analyzer OCMA 500 Instruction Manual CODE GZ0000331784D ...
Страница 20: ......
Страница 58: ...Measurement 38 Flow scheme of operation Fig 36 Operation flow of automatic measurement ...
Страница 65: ...Measurement 45 Manual measurement Flow scheme of operation Fig 45 Operation flow of manual measurement ...