Continuous judgment time setting
This screen enables to set the judgment time of oil film alarm or oil film warning.
Fig. 30 Continuous judgment time setting screen
Operation
The setting value can be changed by each digit.
The following procedure shows the example in which the continuous judgment time (warning)
is set to 120 seconds:
1.
Touch the digit of 100 for “WARNING”.
The background color of the numerical value turns to white.
2.
Turn the value to “1” by
/
keys.
3.
Touch the digit of 10 for “WARNING”.
The background color of the numerical value turns to white (the background color of the
digit of 100 is blue).
4.
Turn the value to “2” by
/
keys.
5.
Touch the digit of 1 for “WARNING”.
The background color of the numerical value turns to white (the background color of the
digit of 10 is blue).
6.
Turn the value to “0” by
/
keys.
7.
Touch “ENT”.
The screen returns to Menu.
To change also the continuous judgment time (alarm), touch the digit of 100 for “ALARM” after
the step of 6. above, and repeat the steps of 1. to 7. above.
Continuous judgment
time (warning)
Continuous judgment
time (alarm)
Name
Description
Continuous judgment
time (warning)
Setting range: 0 seconds to 999 seconds. For the setting step, refer to “Operation” below.
Continuous judgment
time (alarm)
Setting range: 0 seconds to 999 seconds. For the setting step, refer to “Operation” below.