LAQUAtwin series (EC) Operation

2

CODE: GZ0000472457

Basic Operation

Calibration

Calibration is required before measurement.
Use standard solution within the measurement
range in the specifications. 

Tip

Calibration values are saved even if the meter
is switched OFF.

Calibration value is rewritten if calibration is
repeated using the same standard solution.

Measurement

Note

If a measured value is out of the specified
measurement range, "Or" is displayed for
upper range and "Ur" is displayed for under
range.

Ambient air may cause the measured values
to fluctuate. To reduce environmental interfer-
ence, close the protection cover.

When you have a problem with the calibration
or measurement, refer to frequently asked
questions.

Measurement display change

Measurement display change is available on
LAQUAtwin-EC-22 and LAQUAtwin-EC-33.
The display mode switches as follows by press-
ing the MEAS switch in the AS mode.

LAQUAtwin-EC-22:
Between conductivity and temperature alter-
nately

LAQUAtwin-pH-33:
Among conductivity, TDS, and temperature

Power ON

1. Press and hold the ON/

OFF switch.

The power is switched ON,
and the meter model num-
ber is displayed on the LCD.

Power OFF

1. Press and hold the ON/OFF switch.

The power is switched OFF.

Precaution on sample setting 

Place an appropriate amount of a sample or stan-
dard solution into the measurement cell without
trapping bubbles inside. If not, the measurement
may be inaccurate.

Calibration points

The number of calibration points is dependent on
the meter model.

LAQUAtwin-EC-11: 
Up to two-point calibration at 1413 

μ

S/cm and

12.88 mS/cm

LAQUAtwin-EC-22 and LAQUAtwin-EC-33:
Up to three-point calibration at 1413 

μ

S/cm,

12.88 mS/cm, and 111.8 mS/cm

Appropriate

Too little

Cell Electrode

Cross-section of the measurement cell

Multi-point calibration

1. Open the protection cover and place some

drops of the standard solution into the
measurement cell. 

Rinsing the sensor with the standard solution
beforehand will provide a more accurate cali-
bration as it will reduce sample crossover con-
tamination.

2. Close the protection cover and press the

CAL switch.

 and 

 blink, and the calibration value is

displayed.

After the calibration is complete, 

 and 

stop blinking and the measured value is dis-
played. 
The calibration value at 25

C is displayed for 1

s and the display returns to the measurement
mode automatically.

3. Open the protection cover and remove the

standard solution. Then remove moisture
on the sensor by gently dabbing with a soft
tissue.

This completes the 1st point calibration.

4. To perform 2nd point calibration, repeat

steps 1. to 3.

Calibration error

If 

 blinks and Er4 (error dis-

play) appears, the calibration
has failed. 
Perform electrode conditioning. 
Check that the correct standard solution is used,
and repeat calibration after cleaning the sensor.
If the calibration repeatedly fails when using the
correct standard solution(s), the sensor may
have deteriorated. Replace the sensor with new
one.

Protection cover

Sample setting

1. Open the protection cover and put some

drops of sample into the measurement cell.

2. Close the protection cover.

Measurement mode

The auto stable (AS) mode and the auto hold
(AH) mode can be selected. Refer to " 

 Mea-

surement mode change" (page 4) for the opera-
tion to set the measurement mode.

Auto stable (AS) mode

This is the default setting. 

 appears when

the measured value meets the stability criteria.
If the value changes, 

 disappears.

1. Confirm that the meter is in the measure-

ment mode, and place a sample on the sen-
sor.

When the read value meets
the stability criteria, 
appears and the reading is
locked.

2. Document the displayed value when 

appears.

If the read value does not meet the stability cri-
teria, 

 disappears and the reading changes

with time.

Auto hold (AH) mode

 appears when the measured value meets

the stability criteria. The reading then locks
and will not change until the MEAS switch is
pressed for the next measurement.

1. Confirm that the meter is in the measure-

ment mode, and place a sample on the sen-
sor.

2. Press the MEAS switch.

The auto hold function is acti-
vated. 

 blinks until the mea-

sured value has stabilized.

When the measured value is stable, 

 stops

blinking and the displayed value is locked with

 and 

 displayed simultaneously.

3. Document the displayed value.
4. 
Press the MEAS switch.

The auto hold function is deactivated and 
disappears.
Be sure to perform this step before starting the
next measurement. Or, you may mistake the
displayed hold value for the next measured
value.