![horiba LAQUAact ES-71G Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/horiba/laquaact-es-71g/laquaact-es-71g_instruction-manual_2162258064.webp)
Performing test printing of the printer unit
44
●
Performing test printing of the printer unit
In order to check whether the printer unit is operating correctly or there is a printer
communication problem, you can perform test printing.
Connect the instrument and a printer correctly and perform the following procedure for
test printing. As a result of test printing, if the printout is as shown below, the printer unit
is operating correctly.
1. Press the
key to enter the setting mode.
2. Press the
keys to select the "MAINT" (maintenance setting) and then press
the
key.
3. Press the
keys to select the "PRINT" (test print) and then press the
key.
Printing starts automatically. The printer icon blinks during printing. When printing
ends, the printer icon lights and "END" appears. Press the
key.
To return to the setting mode, press the
key.
key
key
key
Printed