pH Setup
19
■
Setup
This section describes all the setup functions available in LAQUA PH1500 series handheld
meters.
● P1 pH Setup
Using P1 pH setup function of the meter, you can:
• Select buffer standard
• Set calibration alarm
• Erase calibration data
To set the pH functions, follow the procedure detailed below:
Prerequisites
Switch on the pH meter.
• Default buffer setup is
BUFF USA
. You can change it to
BUFF NIST
or
BUFF DIN
if
required.
• Calibration alarm setup option must be used to avoid “Calibration interval alarm error” on
page 37 . You can set the calibration alarm for ---- day to upto 90 days, where ----
indicates “no calibration alarm” has been set.
• Erasing previous calibration data is recommended for accurate calibration. Default setup is
NO
but to erase the calibration data, you need to change the setup to
YES
.
To return to the measurement mode, press the
key.
Note
Tip