ABX Micros
ES60
OT/CT
4–4
User Manual - RAB237AEN
◆
Then, the instrument will perform a blank cycle for a reference blank count (an analysis cycle
based on reagents without any blood sample).
A dialog box is displayed to indicate the startup cycle progress. This dialog box also contains an
"
Emergency
stop
" button to stop the startup cycle if necessary.
◆
The instrument displays and prints out the blank cycle results if this function is activated (See
Section 6,
6.5. Analysis running, page 6-14
). This function is activated by default.
Reference blank limits:
1-
Check and verify that the reference blank counts do not exceed the following parameter limits:
WBC < 0.3x10
3
/mm
3
RBC < 0.02x10
6
/mm
3
HGB < 0.3g/dl
PLT < 10x10
3
/mm
3
2-
Press "
OK
" button to validate blank results.
The user identification screen is displayed (See
1.4. User identification, page 4-4
◆
If the reference blank count is above any of the parameter limits, the ABX Micros
ES60
OT/CT will
automatically perform another startup cycle. The cycle progress bargraph restarts at the
beginning. If the problem persists after 3 consecutive cycles, a message will be displayed stating
"
Startup cycle failed
for identifying and resolving
the problem.
1.4. User identification
1.4.1. User profile selection
◆
Select user profile by pressing the appropriate icon (
- The "
Lab tech
" corresponds to the laboratory technician who runs the analyses.
- The "
Scientist
" correponds to the person in charge of the laboratory. He manages the system.
◆
The following table gives the function available according to the user profile selected.
Tab.4-1 User profiles
If any analysis cycles are run prior to a startup cycle, the following message
is displayed: "
Do you want to execute a startup cycle before analysis?
".
Press "
OK
" button to execute the startup cycle or "
Cancel
" button to continue
without startup cycle.
When the ABX Micros
ES60
OT/CT has not been used within 4 hours (default
value, to set it: See Section 6,
8.2.1. Settings startup cycle, page 6-17
)of the
last cycle, the following error message is displayed: "
4 hours elapsed since
last analysis
". It is necessary to perform a startup cycle before running an
analysis cycle.
The "
Technician
" profile is dedicated to the HORIBA ABX local representative
technician only.
User actions
Lab tech
Scientist
Technician
Instrument startup and shutdown
X
X
X
Displaying alarms in "
Status
" area
X
X
X
Running patient analysis
X
X
X
Running control analysis
X
X
X
Displaying and printing patient results
(current and history)
X
X
X
Sending patient results
X
X
X
Содержание ABX MICROS ES60
Страница 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Страница 2: ......
Страница 6: ...User Manual RAB237AEN...
Страница 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Страница 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Страница 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Страница 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Страница 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...