ABX Micros
ES60
OT/CT
7–38
User Manual - RAB237AEN
6.1.2. QC alarms
6.1.3. Reagent alarms
Diag.7-43 QC status button
System alarm
message
Help message
Possible cause / Corrective action
Invalid QC "lot
number"
Check the last QC "lot number" result
in QC menu and rerun the appropriate
control if necessary.
A QC analysis is out of normality limits:
- Check the last control results.
- Run an analysis for this control (See Section 4,
Running control blood (ABX Minotrol 16), page 4-11
).
- Replace the control lot (See Section 5,
).
Inter QC time limit
"lot number"
Inter QC "lot number" time set
elapsed. You should run the
appropriate control.
The inter QC time limit set for a control is elapsed:
- Run an analysis for this control (See Section 4,
Running control blood (ABX Minotrol 16), page 4-11
).
XB drift
Check XB data in QC menu/XB screen.
A statistic dot is out of limits:
- Check XB data.
Diag.7-44 Reagent status button
System alarm
message
Help message
Possible cause / Corrective action
Reagent LYSE expired.
Check LYSE expiration date in Maintenance menu
/ Reagent screen. If necessary, replace the
appropriate reagent by another lot number, run
priming cycle from reagent screen and run a
Startup cycle.
- Replace the LYSE reagent.
See
4.2. Reagent management, page 7-21
Reagent DILUENT
expired.
Check DILUENT expiration date in Maintenance
menu / Reagent screen. If necessary, replace the
appropriate reagent by another lot number, run
priming cycle from reagent screen and run a
Startup cycle.
- Replace the DILUENT reagent.
See
4.2. Reagent management, page 7-21
Reagent CLEANER
expired.
Check CLEANER expiration date in Maintenance
menu / Reagent screen. If necessary, replace the
appropriate reagent by another lot number, run
priming cycle from reagent screen and run a
Startup cycle.
- Replace the CLEANER reagent.
See
4.2. Reagent management, page 7-21
Reagent PACK
expired.
Check PACK expiration date in Maintenance menu
/ Reagent screen. If necessary, replace the
appropriate reagent by another lot number, run
priming cycle from reagent screen and run a
Startup cycle.
- Replace the reagent pack.
See
4.2. Reagent management, page 7-21
Low reagent level
LYSE
Check LYSE level in Maintenance menu / Reagent
screen. If necessary, replace the appropriate
reagent, run priming cycle from reagent screen
and run a Startup cycle.
LYSE reagent is almost empty (counter of
tests lower than 10):
- Replace the LYSE reagent.
See
Содержание ABX MICROS ES60
Страница 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Страница 2: ......
Страница 6: ...User Manual RAB237AEN...
Страница 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Страница 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Страница 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Страница 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Страница 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...