Conductivity electrode (9383-10D)
Instruction Manual
Thank you for purchasing the HORIBA electrode. Read this
manual carefully before using the electrode.
Conformable standards
Authorised representative in EU
HORIBA UK Limited
Kyoto Close Moulton Park Northampton NN3 6FL UK
Cautions during handling
Do not allow the electrode to come in contact with any
hard surface.
If the electrode pole plate is dry, immerse the electrode
in pure (ion exchange) water for at least one hour, prior
to use.
The electrode connector requires high insulation. Do not
allow the connector to be in contact with water or dirty
hands.
Do not rub the electrode plate with a brush or polish it
with a polishing agent.
The water-resistant construction of meters can be used
in combination with this electrode to provide water-
resistant construction (conforming to IP-67). When
measuring, do not immerse the cap, the lead or the
connector in the sample.
Packaged contents
Specifications and names of parts
Names of parts
Specifications
This equipment conforms to the following standards:
RoHS:
EN50581
9. Monitoring and control instruments
Name
Q’ty
Electrode
1 pc
Operation Manual
1 copy
Model
9382-10D
Cell constant
100 m
-1
(Former units: 1 cm
-1
)
Measurement range
0.1 mS/m to 10 S/m (Former units: 1
μ
S/cm to
100 mS/cm)
Usable temp. range
0
°
C to 80
°
C
Storage temp. range
0
°
C to 50
°
C
Electrode material
Titanium - platinum black
Wetted part materials
PPS, PSF and titanium
Electrode
connector
Connector cover
Cap
Lead
Temp. connector (T electrode)
Protective cap
Electrode plate
Preparations
Connecting to meter
1. Insert the electrode connector into the connector port
sleeve on the meter, after aligning with the pin. Do not
insert the connector unless it is aligned properly with the
connector port.
2. Press the electrode connector into the connector port on
the meter, while turning the connector to the right.
3. Slide the connector cover over the connector. Then,
push the cover in straight until it comes in light contact
with the meter case.
Do not turn the cover.
4. Insert the temperature connector into the jack on the
meter. Insert the connector firmly, until the O-ring on the
connector can no longer be seen.
Setting cell constant
The cell constant for this electrode is displayed on the
electrode cap.
Enter the cell constant (shown as 0.954 x 100, in this
example) of the electrode, by following the instructions in the
Operation Manual for the meter.
Example:
9383-10D
LOT
. ......
0.954 x 100 m
-1
Preparing electrode
Either wash the electrode plate using a bottle that contains
pure (ion exchange) water, or immerse the electrode in a
beaker containing pure (ion exchange) water and lift the
electrode up and down a few times to rinse it, then wipe it
dry using filter or tissue paper.
Electrode length
150 mm (incl. cap)
Max. height of electrode
plate
53 mm (position from electrode tip)
Ext. diam. of wetted part
16 mm
Lead length
1 m
C1
T1
T2
C2
①
②
③
④