Optical DO Sensor
300-D-2 / 300-D-5
*”DO” is an abbreviation of Dissolved Oxygen.
Warranty and Responsibility
HORIBA Advanced Techno Co., Ltd. warrants DO sensor shall be
free from defects in material and workmanship and agrees to repair
or replace free of charge, at option of HORIBA Advanced Techno
Co., Ltd., any malfunctioned or damaged product attributable to
responsibility of HORIBA Advanced Techno Co., Ltd. for a period of
Three (3) years from the delivery unless otherwise agreed in a
written statement. Sensor cap is eligible for a replacement against
incipient defects within Two (2) years from the delivery. In any one of
the following cases, none of the warranties set forth herein shall be
extended:
Any malfunction or damage attributable to improper operation
Any malfunction attributable to repair or modification by any
person not authorized by HORIBA Advanced Techno Co., Ltd.
Any malfunction or damage attributable to the use in an
environment not specified in this manual
Any malfunction or damage attributable to violation of the
instructions in this manual or operations in the manner not
specified in this manual
Any malfunction or damage attributable to any cause or causes
beyond the reasonable control of HORIBA Advanced Techno
Co., Ltd. such as natural disasters
Any deterioration in appearance attributable to corrosion, rust,
and so on
HORIBA Advanced Techno Co., Ltd. SHALL NOT BE LIABLE FOR
ANY DAMAGES RESULTING FROM ANY MALFUNCTIONS OF
THE PRODUCT, ANY ERASURE OF DATA, OR ANY OTHER USES
OF THE PRODUCT.
Conformable standards
This equipment conforms to the following standards:
Authorised representative in EU
HORIBA Europe GmbH
Hans-Mess-Str.6, D-61440 Oberursel, Germany
Cautions during handling
Do not allow the sensor to be in contact with any hard surface.
The surfaces of the sensor cap is covered with
fluorescent materials. Do not allow the sensor cap to come in
contact with dirt and operate carefully when immersing into a
solution.
Avoid continuous measurements in solutions with strong acids,
solutions with strong alkalis and solutions with organic solvents.
Electromagnetic interferences may cause malfunctions in the
DO sensor.
Packaged contents
Specifications and names of parts
Specifications
RoHS:
EN50581
9. Monitoring and control instruments
DO sensor (with sensor cap)
1 pc.
Calibration bottle
1 pc.
Protective guard
1 pc.
Instruction manual
1 copy
Model
300-D-2/300-D-5
Measurement method
Fluorescence quenching
Measurement range
0 to 20 mg/L
Accuracy (DO)
±2 % *
Temperature range
0 to 50 °C
Accuracy (Temperature)
±0.3 °C
Response time
30 s (t :90)
Watted materials
DO sensor: PC/ABS
Sensor cap: PMMA
Protective guard: SUS316L
DO sensor size
16×200
Lead cable length
2 m (300-D-2)
5 m (300-D-5)
Waterproof
IP67
*For a calibration perfomed in saturated water
Names of parts
Note
The sensor cap is delivered equiped with the DO sensor.
Perform a calibration before any measurement.
The calibration method is indicated in the instruction
manual of the meter.
It is recommended to make a calibration in liquids for a
better accuracy during measurements.
Make sure there are no air bubbles visible on the surface of
the cap during a measurement.
After finishing a measurement, clean the sensor by using a
soft cloth. Take care not to scratch the sensor cap. After
cleaning, rinse the sensor cap with distilled water.
Store the DO sensor with the sensor cap and the calibration
bottle in a cool, dark location. The sponge inside the
calibration bottle should be kept wetted.
Sensor cap replacement instruction procedure
1. Remove the old sensor cap from the DO sensor.
2. Tighten the new sensor cap on the DO sensor.
3. Input the membrane parameter indicated on the sensor cap in
the meter.
4. Perform calibration.
The instruction procedure has been completed.
Note
The membrane parameter of the sensor cap delivered is set
in the meter.
The setting method of the membrane parameter is indicated
in the instruction manual of the meter.
DO sensor
Calibration bottle
Sensor cap
Protective cap
(300-D-M)
DO sensor
Sensor cap
xxxx
Sensor cap
DO sensor
Membrane parameter
Enlarged